![Yamato FO 100 Скачать руководство пользователя страница 37](http://html.mh-extra.com/html/yamato/fo-100/fo-100_instruction-manual_905271037.webp)
35
In the Event of Failure…
Trouble Shooting
Problem Possible
Cause
Solution
Bad condition of earth leakage breaker
Bad condition of power switch
Replace the part
The device does not start
when turning on the power
switch.
Problem in power source
Use an appropriate power supply.
Heater is disconnected
Thermal fuse is disconnected
Temperature does not rise.
Bad condition of thermoregulator
Replace the part
Bad condition of temperature sensor
Heater is disconnected
Bad condition of thermoregulator
Replace the part
Takes too much time to
raise temperature.
Problem in power source
Use an appropriate power supply.
Bad condition of SSR
Heater is not turned off after
reaching to preset value.
Bad condition of thermoregulator
Replace the part
Bad condition of fan
Thermal fuse blows out.
Bad condition of thermoregulator
Replace the part
When power failure occurs…
•
When power is supplied after a power failure, the device automatically starts operation again with
the same state as just before the power failure. It is danger that the device starts unattached
operation after a power failure.
•
We recommend for you to turn off the switch of device if a power failure occurs during operation.
In the case if the error other than listed above occurred, turn off the power switch and primary
power source immediately. Contact the shop of your purchase or nearest Yamato Scientific