background image

YCO-1102-05-02

CPS Series

Operation Manual

III Setup

Installing the Battery

The CPS uses two standard AAA alkaline batteries.  The battery compartment is located under the

scale towards the back of the scale.  Press the cover in and out to open.  Place the batteries into
the battery compartment making sure they are aligned correctly with the polarity markings inside the
compartment.  Reversed polarity may damage the scale.  Replace the compartment cover.

Setting the Scale Functions

The scale functions can be changed in the setup mode to select which units are displayed (g, oz, dwt

and ozt), to select whether or not the counts are displayed, to change the duration of the auto-off
function, to disable the auto-off function and to change the capacity of the scale when the load cell
and panel are changed.

To enter the setup mode, press the 

Tare

 key for about three seconds.  The display will briefly show

“SET” followed by “UNIT”.  Use the 

Count

 key as an ENTER or RETURN key, and use the 

Unit

 key

as a SELECT or TOGGLE key.  Refer to the setup chart on the following page.  Any unit that is
selected “ON” will be displayed, in sequence, when the 

Unit

 key is pressed.  The auto-off setting

selects the duration, in seconds, after which the scale will automatically turn itself off if it is not being
used.  If “OFF” is selected, the auto-off function will be disabled.  The capacity setting should only
be changed if the load cell and front panel are being changed.

Calibrating the Scale

With the scale on a level and stable surface, power the scale on and, with nothing on the platform,

press the 

Unit

 key for about three seconds.  The display indicate “CAL” followed by “CAL 0”.  The

scale will calibrate the zero load point.  The display will then alternate “LoAd” and “100 g”.  Place
100 g on the scale platform.  After a few seconds, the scale will indicate “donE” and return to nor-
mal weighing mode.  Remove the calibration weights from the scale platform and test the scale
calibration.

If the calibration is inaccurate, verify that the surface the scale is on is level, stable and free from

vibration.  Recalibrate the scale.  If the scale still will not read accurately, service may be required.

2

Содержание CPS series

Страница 1: ...ORPORATION 1775 S Murray Blvd Colorado Springs CO 80916 USA Tel 719 591 1500 Fax 719 591 1045 YAMATO TECH CORPORATION 112 19425 Langley By Pass Surrey B C V3S 6K1 Canada Tel 604 533 2338 Fax 604 533 0827 www yamatocorp com Pocket Scale Counting Feature ...

Страница 2: ...o the scale 1 Do not push the indicator or keys with sharp objects They may break or puncture the switch membrane panel 2 Use the specified power supply and choose a suitable environment If you do not the weight readings may be inaccurate and the scale may be damaged 3 The scale is a sensitive weighing instrument avoid physical shocks If you drop something on the scale step on the platform or drop...

Страница 3: ...p mode 6 Count key Press to enter count mode and to select sample size 10 or 20 1 Weighing system Strain gauge load cell Platform 2 4 x 3 0 60 x 75 mm ABS polymer Dimensions 5 2 D x 3 2 W x 0 9 H 133 x 80 x 23 mm Weight display Single face display Type 7 segment LCD Characters 0 5 H x 0 15 W 12 5 x 4 mm 5 digits Power supply 2 AAA alkaline batteries Battery life 14 hours of continuous use Operatin...

Страница 4: ...he Unit key as a SELECT or TOGGLE key Refer to the setup chart on the following page Any unit that is selected ON will be displayed in sequence when the Unit key is pressed The auto off setting selects the duration in seconds after which the scale will automatically turn itself off if it is not being used If OFF is selected the auto off function will be disabled The capacity setting should only be...

Страница 5: ...ess for three seconds Return to normal weighing mode Tare Unit Unit Unit Unit Unit Unit Unit Unit Unit Unit Unit Unit Unit Unit Unit Unit Unit Unit Unit Unit Unit Count Count Count Count Count Count Count Count Count g g g oz dwt ozt g oz ozt dwt g oz dwt ozt ...

Страница 6: ...the Scale Functions to change this feature Low Battery Condition The low battery indication will appear when the battery needs to be replaced See section III Installing the Battery to replace the battery Zero the Display The scale will automatically zero when turned on If a value is displayed when the platform is empty push the Tare key to zero the display The maximum zero range is approximately t...

Страница 7: ... Place the first item in the container until the desired weight is reached 4 Press the Tare key and the display will return to zero Place the next item in the container until the desired weight is reached 5 Continue in this manner until all items to be batched have been weighed Do not allow the gross weight on the scale platform to exceed the capacity of the scale 6 To return to gross weighing mod...

Страница 8: ... or service is required Power on counts are too high Remove weight from the platform turn the scale off and back on If the error remains calibration or service is required Underload condition Generally indicates that something is wedged between the platform and scale body Remove the obstruction If there is no obstruction calibration or service is required Power on counts are too low This generally...

Отзывы: