background image

YCO-1093-04-02

CPC-2000

Operation Manual

The scale can check various functions automatically. When a problem is detected, the following mes-

sages will be displayed. If you take the steps described below but the problem is not fixed, the scale
needs to be serviced by a qualified scale technician.

Low battery indication.  Connect the AC adaptor to the scale
and recharge the battery.  This takes approximately 14
hours.  The scale can be used while it is recharging.  If the
battery does not charge, see a qualified scale technician.

Power on counts are greater than calibrated zero plus 10%.

Remove weight from the platform and press the 

ON/OFF

 key, if

necessary.  If no weight is on the platform, calibration or
service is required.

Power on counts are less than calibrated zero minus 10%.
This generally indicates that the platform is not on the scale
or that something is wedged between the platform and the
scale body.  Replace the scale platform, remove any obstruc-

tions under the platform and press the 

ON/OFF

 key, if neces-

sary.  Otherwise, calibration or service is required.

Overload condition.  Remove weight that is greater than the
scale capacity from the platform.  If no weight is on the
platform, calibration or service is required.

Underload condition.  Generally indicates that something is
wedged between the platform and scale body.  Remove the
obstruction.  If there is no obstruction, calibration or service is
required.

Calibration span is too high.  Calibration or service is re-
quired.

EEPROM R/W error.  Service is required.

V Troubleshooting and Error Messages

6

Содержание CPC-2000

Страница 1: ...4 02 YAMATO CORPORATION 1775 S Murray Blvd Colorado Springs CO 80916 USA Tel 719 591 1500 Fax 719 591 1045 YAMATO TECH CORPORATION 112 19425 Langley By Pass Surrey B C V3S 6K1 Canada Tel 604 533 2338...

Страница 2: ...acy 4 If the scale becomes dirty wipe it with a soft cloth For stubborn stains apply a little neutral detergent and then wipe the scale with a dry cloth Do not use thinner benzene hot water or chemica...

Страница 3: ...for five seconds to turn scale off 7 Net Gross key Toggles between net and gross weight display 8 Tare key Tares off the weight on the scale platform 1 UNIT TARE RECHARGE WEIGHT N kg lb G oz ON OFF N...

Страница 4: ...If the scale is not level the readings may not be accurate Setting the User Functions The scale keywords can be set in the user setup mode to change the duration of the auto off function to disable th...

Страница 5: ...02 CPC 2000 Operation Manual III Setup 3 UNIT UNIT UNIT UNIT UNIT UNIT UNIT UNIT UNIT UNIT UNIT UNIT UNIT UNIT N G N G N G N G N G N G Press for three seconds Return to normal operating mode kg lb oz...

Страница 6: ...The low battery indicator will appear when the battery pack needs to be recharged Insert the AC adaptor into the jack on the back of the scale The scale may be operated while it is recharging Zero th...

Страница 7: ...To return to gross zero remove all items from the scale platform and press the N G key 6 To return to gross weighing mode remove all items from the scale platform and press the TARE key Batching 1 Beg...

Страница 8: ...platform calibration or service is required Power on counts are less than calibrated zero minus 10 This generally indicates that the platform is not on the scale or that something is wedged between th...

Отзывы: