CALIBRATION PROCEDURE
Operation
Display
1. Place the scale on a hard, level
surface.
2. Don’t press ON/OFF key,
3. Press and hold ON/OFF, and then
press UNIT key three times. To enter
into calibration procedure.
4. Display will show the Count value
(256 ~ 250000).
5. Then press the
key.
6. Place the calibration weight (10kg)
carefully in the centre of the scale
plate, and the display will show SPAN
Count value (4000-80000).
7. Press the
key. The
adjustment process is started.
8. When the adjustment is finished
successfully, the scale is
automatically switched over to the
weighing mode again and the
calibration weight value will be
displayed.
9.
Remove the calibration weight.
Re-power on, the instrument is in
normal mode.
11
INSTRUCTION AUTO SHUT-OFF FUNCTION
Define
Select
Press UNIT