60
Maintenance and Storage
Depth Sensor Maintenance and Storage
Depth Sensor Storage
The ProDSS optional depth sensor on 4 port ProDSS cables accesses
water through ports located in the bulkhead (
Figure 63
). Although not
directly accessible, correct maintenance and storage is necessary for
reliable operation.
The depth sensor can be stored dry, in water-saturated air or submerged
in water.
NOTICE:
To prevent damage to the sensor’s strain gauge, do not
store the sensor in corrosive solutions.
Depth Sensor Maintenance
Periodically clean the depth ports with the syringe included in the ProDSS
maintenance kit (626990). Fill the syringe with clean water and gently
force water into one of the ports. Flush until clean water flows from the
opposite depth port.
NOTICE:
Do not insert objects into the depth ports. Damage to
the depth transducer from incorrect cleaning is not
covered by the warranty.
Figure 63
Depth sensor flush
Содержание YSI ProDSS