To adjust the salinity compensation value:.
1.
Use the up or down arrow keys to highlight the salinity box on the run screen,
then press enter.
2.
Use the up or down arrow keys to adjust the salinity compensation value to the
salinity of the water you are testing.
3.
Press enter to confirm and to save the new salinity compensation value.
Taking Measurements
The Pro20/Pro20i can be calibrated in
%
saturation,
mg/L
, or
ppm
. Calibration of any
option (%, mg/L, or ppm) will automatically calibrate the others. The Pro20/Pro20i can
be calibrated with the press of one key when One Touch Cal is enabled in the System
Setup menu.
The following procedure outlines the % saturation calibration option with and without
One Touch Cal enabled.
1.
Turn the instrument on and wait 5-15 minutes if using a polarographic sensor.
2.
To take readings, insert the probe into the sample. Move the probe in the
sample at a rate of at least 6 inches per second until the readings stabilize.
3.
Highlight
Save
and press
Enter
to store the reading. The instrument will
confirm that the reading was successfully saved.