background image

8

9

Figure 7, page 1 of System Setup menu. 

aUdIo  

Audio can be enabled by highlighting Audio and pressing enter. When 
enabled, there will be an ‘X’ in the box next to Audio. 

When Audio is enabled, the Pro1030 will beep twice to indicate stability 
when Auto Stable is enabled. The instrument will also beep when a key 
is pressed.  When Audio is disabled, the Pro1030 will not beep. 

contrast

To adjust the display Contrast, use the up or down arrow keys to 
highlight Contrast, then press enter. Next, use the up or down arrow 
keys to adjust the contrast.  The up arrow key will darken the contrast 
and the down arrow key will lighten the contrast. After adjusting the 
contrast, press enter to save and exit the Contrast adjustment function.   

ALTERNATE CoNTRAST ADJUSTMENT oPTIoN

If necessary, there is an alternate method of adjusting the contrast. To 
adjust the contrast, press and hold the menu key, then press the up 
arrow key to darken the contrast or press the down arrow key to lighten 
the contrast.  

teMperatUre UnIts

Highlight Temperature Units and press enter to open a submenu that 
allows you to change the temperature units displayed on the Run 

screen. Highlight the desired unit (Celsius or Fahrenheit) and press 
enter to enable. The enabled temperature unit will have an ‘X’ in the 
box next to it. only one unit may be enabled at a time. Highlight the 
ESC-Exit box and press enter to save any changes and to close the 
Temperature Units submenu.

Ise sensor type

ISE Sensor Type sets the type of ISE sensor being used; either pH  
(model #1001) or oRP (model #1002).
  
Use the up or down arrow keys to highlight ISE Sensor Type, then press 
enter to open a submenu. Highlight the sensor type corresponding 
to the sensor installed on the cable and press enter to confirm.  The 
enabled sensor type will have an ‘X’ in the box next to it. Next, use the 
down arrow key to highlight the ESC – Exit, then press enter to save 
changes and to close the sensor submenu.  

Ise UnIts

Highlight ISE Units and press enter to open a submenu that allows you 
to select the ISE units to be displayed on the Run screen. Highlight a 
unit and press enter to enable or disable it. An enabled ISE unit will 
have an ‘X’ in the box next to it. Highlight the ESC-Exit box along the 
bottom of the display and press enter to save any changes and to close 
the ISE Units submenu. 

When pH is enabled in the ISE Sensor Type menu, there are two 
selectable measurement units: pH and pH mV.  pH mV is the sensor’s 
electrical measurement signal before being converting into pH 
units. pH mVs can help you determine if you are performing a good 
calibration and the condition of the pH electrode. 

When oRP is enabled in the ISE Sensor Type menu, only oRP mVs can 
be enabled as the ISE unit.  

aUto staBle

Auto Stable utilizes preset values to indicate when a reading is stable. 
The preset values are adjustable in the System Setup menu. The user 
can input a % change in measurement reading over ‘x’ amount of time 
in seconds. There are two separate Auto Stable controls, one for ISE 
readings (ISE Auto Stable) and one for conductivity readings (Cond. 
Auto Stable).  ISE Auto Stable is located on the first page of the System 
Setup menu. Cond. Auto Stable is located on the second page of the 
System Setup menu. 

Содержание YSI Pro1030

Страница 1: ...USER MANUAL English Pro1030...

Страница 2: ...First Power On 7 System Setup Menu 7 Audio 8 Contrast 8 Temperature Units 8 ISE Sensor Type 9 ISE Units 9 Auto Stable 9 pH Buffer Set 10 Conductivity Units Cond Units 10 Specific Conductance Reference...

Страница 3: ...red by this warranty To exercise this warranty call your local YSI representative or contact YSI Customer Service in Yellow Springs Ohio at 1 937 767 7241 800 897 4151 or visit www YSI com for a Produ...

Страница 4: ...pack the instrument and accessories and inspect for damage Compare received parts with items on the packing list If any parts or materials are damaged or missing contact YSI Customer Service at 800 89...

Страница 5: ...u from the Run screen 2 Up Arrow Use to navigate through menus to navigate through box options along the bottom of the Run screen and to increase numerical inputs 3 Power and Backlight Press once to t...

Страница 6: ...y 1 o ring visible Failure to properly seat the sensor may result in damage 6 Twist the sensor clockwise to engage the threads and finger tighten Do NOT use a tool This connection is water tight The I...

Страница 7: ...System Setup menu the up and down arrow keys will move the highlighted bar up and down the system setup options See the System Setup menu section of this manual for more information about these option...

Страница 8: ...ther pH model 1001 or ORP model 1002 Use the up or down arrow keys to highlight ISE Sensor Type then press enter to open a submenu Highlight the sensor type corresponding to the sensor installed on th...

Страница 9: ...enter to enable or disable it An enabled conductivity unit will have an X in the box next to it Highlight the ESC Exit box along the bottom of the display and press enter to save any changes and to cl...

Страница 10: ...e a multiplier for the water at your sampling site Use the following procedure to determine the multiplier for a specific sample 1 Determine the specific conductance of a water sample from the site 2...

Страница 11: ...e and 10 seconds pH Buffer Set USA Conductivity Units cond mS cm and spc mS cm Parameter Reset Defaults Conductivity Auto Stable Off 0 0 Change and 10 seconds SPC Reference Temperature 25 C SPC Temper...

Страница 12: ...hold Cal for three seconds 3 Highlight pH and press enter If pH is not listed as an option check the System Setup menu to ensure pH is enabled in the ISE Sensor Type menu 4 Highlight 3 point and pres...

Страница 13: ...hloride when calibrating Specific Conductance since these solutions typically have a Temperature Coefficient of 1 91 C Additionally be sure to enter the value of the solution as it is listed for 25 C...

Страница 14: ...ess the Cal key to exit the calibration error message and return to the Run screen See the Troubleshooting section of this manual for possible solutions Taking M easurements Before taking measurements...

Страница 15: ...the data set number 01 through 50 Erasing Data While viewing saved data press the enter key to access the function boxes at the bottom of the display Next use the up or down arrow keys to highlight E...

Страница 16: ...prevent water from entering the port Once the ISE sensor is removed examine the connector inside the port If any moisture is present use compressed air to completely dry the connector or let it air d...

Страница 17: ...h all parts of the sensor tip without stress You can also use a pipe cleaner for this operation if more convenient i Dry the port and sensor connector with compressed air and apply a very thin coat of...

Страница 18: ...displays Calibration Over Calibration Under or Unstable Reading during calibration 1 Ensure the conductivity sensor is clean Follow the cleaning procedures in the Care Maintenance and Storage section...

Страница 19: ...uctivity 0 500 uS cm 0 5 mS cm 0 50 mS cm 0 200 mS cm auto ranging 0 0001 to 0 1 mS cm 0 1 to 0 uS cm range dependent Instrument only 0 5 of the reading or 1 uS cm whichever is greater Instrument with...

Отзывы: