ORP Calibration
1.
Place the clean sensor in ORP calibration solution. Wait for the ORP and
temperature readings to stabilize.
2.
Press and hold
Cal
for three seconds.
3.
Highlight
ORP
and press
Enter
. If ORP is not listed as an option, check the
System Setup menu to ensure ORP is enabled in the ISE Sensor Type menu.
4.
Use the up and down arrow keys to adjust the ORP calibration solution value.
5.
Wait for the temperature reading to stabilize, then press
Enter
to complete the
calibration or press
Cal
to cancel.
6.
‘
Calibration Successful
’ will display for a few seconds to indicate a successful
calibration and then the instrument will return to the Run screen.
7.
If the calibration is unsuccessful, an error message will display on the screen.
Press the
Cal
key to exit the calibration error message and return to the Run
screen. Refer to the troubleshooting section of the manual for a possible
solution or contact YSI Technical Support.
7.
Rinse the sensor and place it in the second pH buffer (4/4.01 or 10/9.18). If
necessary, use the up and down arrow keys to adjust the pH buffer value.
8.
Wait approximately 30 to 60 seconds for the pH sensor to stabilize and for the
temperature reading to stabilize.
9.
Press
Enter
to either complete the 2-point calibration or to continue with the 3rd
calibration point. Or, press
Cal
to cancel.
10.
If continuing on with a 3rd calibration point, rinse the sensor and place it in the
third pH buffer (4/4.01 or 10/9.18). If necessary, use the up and down arrow keys
to adjust the pH buffer value.
11.
Wait approximately 30 to 60 seconds for the pH sensor to stabilize and for the
temperature reading to stabilize.
12.
Press
Enter
to complete the 3-point calibration. Or, press
Cal
to cancel.
13.
‘
Calibration Successful
’ will display for a few seconds to indicate a successful
calibration and then the instrument will return to the Run screen.
14.
If the calibration is unsuccessful, an error message will display on the screen.
Press the
Cal
key to exit the calibration error message and return to the Run
screen.