IOEM/REM Safey and Installation Guide

44

IOEM/REM Safey and Installation Guide

45

CAUTION:

 Run high and low voltage cables through separate bulkhead and 

conduit.

5. 

Use (-) and (+) pin for selected input channels 1-4 for IOEM-4 and 1-8 
for IOEM-8 to terminate wires. Terminate the ground wire to the (-) loca-
tion. The 24V pin is used only when wiring a 4/20mA nonpowered device 
that requires 24V loop power from the IOEM to power the 4/20 device.

6. 

Please see individual diagrams for wiring IOEM input devices:

A.  digital device, page 45;
B.  0-1 and 0-5 volt device, page 45;
C.  4/20mA (powered/sink) and (not powered/source) devices, page 46;
D.  temperature sensor instructions (no diagram provided), page 46

CAUTION

:  Inputs are not isolated.  Devices connected must be electrically 

isolated from ground and the water.

 

IOEM input and 4/20mA termination locations

7. 

Complete wiring for all IOEM I/Os that will be used in application.

8. 

Close front panel - page 32.

9. 

Configure 5X00 E-Aux systems using AquaManager PC software. 

Complete E-Aux system configuration information is provided in the 

AquaManager Help system. 

10.  Test E-Aux/IOEM wired systems including display, control, and/or alarm 

functionality based on the particular user-defined configuration.

Installation and Wiring

A. Wire Digital input

Wire Aux Digital Input

• 

Install digital input device according to manufacturer’s instructions.

• 

Use 2 wires to connect Digital input device to an IOEM channel.

• 

Digital devices are mechanical and do not require power.

• 

Connect switch wires to an IOEM I/O channel. There is no polarity for digital 
input device wires; therefore, (+) and (-) terminals are interchangeable.

• 

Configure 5X00 Eaux systems using AquaManager PC software. Complete 

Eaux system configuration information is provided in the AquaManager 

Help system.

Digital input switch

CAUTION: 

Do not input external voltage.

B. Analog  0-1 and 0-5 Volt input

Wire Analog 0-1Volt and 0-5 Volt Input

• 

Install 0-1 and/or 0-5 Volt analog device according to manufacturer’s in-
structions.

• 

Use 2 wires to connect device to an IOEM channel.

• 

Follow polarity from volt source to IOEM channel (+) to (+) and (-) to (-)-
see diagram below.

• 

Configure 5X00 Eaux systems using AquaManager PC software. Complete 

Eaux system configuration information is provided in the AquaManager 

Help system.

Analog 0-1 Volt & 0-5 Volt device

CAUTION: 

 Observe correct polarity on analog input.

Installation and Wiring

Содержание YSI IOEM

Страница 1: ...YSI Incorporated IOEM Input Output REM Relay Expansion Modules Safety and Installation Guide AquaManger configuration quick start guides included at chapter 4...

Страница 2: ...ntal Specifications 12 IOEM Input Output System Specifications 12 REM Output Specifications 13 Component and System Non Input Specifications 13 2 Safety 15 THIS IS IMPORTANT 16 Warnings and Cautions 1...

Страница 3: ...aManager to configure IOEM channel mode 73 Configure 5X00 E Aux system for IOEM input 74 Configure 5X00 system for IOEM 4 20mA output 76 Configure 5X00 system for IOEM PID output control device 77 Res...

Страница 4: ...IOEM REM Safey and Installation Guide 6 1 Introduction IOEM Product Description REM Product Description Product features Using this Manual Specifications This page left intentionally blank...

Страница 5: ...oftware is required and serves as a management tool using either a direct RS232 or TCP IP connection where a 5X00 instrument is configured as a master on the RS485 network 1 32 bit processor networkin...

Страница 6: ...ystems as additional relay outputs on the network REM instruments have 4 relay outputs that are used to connect peripheral devices These peripheral devices are used to control and or alarm networked 5...

Страница 7: ...Directive 2011 65 EU WEEE 2002 96 EC FCC See Appendix 3 Declaration of Conformity page 88 IOEM Input Output System Specifications Input type There are 4 IOEM 4 or 8 IOEM 8 I O channels that allow dig...

Страница 8: ...times per second and readings are averaged Communication Feature Specification RS232 required for firmware upgrading up to 100 ft 30m Instrument to Instrument Networking Feature Specification With a...

Страница 9: ...consignes de s curit ne sont pas suivies Safety CAUTIONS are used when personal injury or damage to equipment can result if safety information is not followed Les notices ATTENTION sont utilis es lors...

Страница 10: ...n plus de la mise la terre du mod le IOEM and REM tous les r servoirs doivent tre mis la terre Safety WARNING Do not overload relays Verify that the surge current at startup does not exceed the relay...

Страница 11: ...le d expansion de relais et sont utilis s comme r f rence flottante au sol Cela exige que les capteurs ne soient pas mis la terre individuellement CAUTION Power supply voltage above 16 5 VDC may perma...

Страница 12: ...IOEM REM Safey and Installation Guide 22 This page left intentionally blank 3 Installation Wiring Installing wiring IOEM and REM components and peripheral devices...

Страница 13: ...ips head screwdriver to remove IOEM and REM front panel to ac cess board assembly necessary tools for cutting mounting holes in mounting brackets rail mount electric drill and bits for mounting IOEM a...

Страница 14: ...REM to the surface Figure 3 2 Installation and Wiring Rail Mount Mount IOEM and REM to handrail or pipe 1 to 1 5 2 5 to 3 8cm in diameter The rail mount kit YSI model number 6509 includes u bolts a pr...

Страница 15: ...al plate is labeled with letters These letters correspond to a label affixed to the inside of the front panel which identifies wire termination Installation and Wiring locations IOEM and REM ground wi...

Страница 16: ...ter so that a watertight seal can be made Rubber grommets can be frozen put in a freezer for 1 hour and then easily drilled to the required diameter Some pre drilled rubber grommets are supplied with...

Страница 17: ...ATTENTION les mod les d IOEM module d expansion d entr et de sortie et de REM module d expansion de relais utilisent des appareils transistoris s sensibles pouvant tre endommag s par des d charges d l...

Страница 18: ...ommended when operating with AC power Backup power can be supplied by 12VDC or UPS battery source Wire AC Power The AC power option includes an integrated universal worldwide AC switching power supply...

Страница 19: ...EM REM Safey and Installation Guide 37 Ferrite bead YSI p n 606399 Ferrite bead assembled CAUTION Run high and low voltage cables through separate bulkhead and conduit Installation and Wiring IOEM AC...

Страница 20: ...d if batteries are to be fully powered when used as back up power source Power supply voltage above 16 5 VDC may permanently damage the IOEM and REM When back up voltage falls below 9 0 volts the IOEM...

Страница 21: ...iring Directions to Wire Relay Outputs WARNING Disconnect external power to the unit before wiring If any of the relay contacts are connected to a hazardous live circuit ALL contacts on ALL relays mus...

Страница 22: ...m E Aux systems are also referred to as virtual inputs 2 as a 4 20mA output for 5X00 sensor systems including pH ORP conductivity temperature DO and aux and E Aux systems The 4 20mA output signal can...

Страница 23: ...10 Test E Aux IOEM wired systems including display control and or alarm functionality based on the particular user defined configuration Installation and Wiring A Wire Digital input Wire Aux Digital I...

Страница 24: ...cables through separate bulkhead and conduit 5 Terminate temperature sensor wires to selected IOEM channel There is no polarity for temperature sensor wires therefore and terminals are interchangeabl...

Страница 25: ...livers this information to the master and the master relays this information to the REM over the RS485 network and relay 4 is energized Installation and Wiring Notes Use an RS485 Line Extender Booster...

Страница 26: ...leanly Nicks stray wires not cleanly terminated can result in unreliable network communication CAUTION Run high and low voltage wires cables through separate bulk head and conduit IOEM network REM net...

Страница 27: ...rips not connected to device When master is first physical device on network as shown above connect terminal strip to master and last slave slave 4 to begin network testing Terminal strips for slaves...

Страница 28: ...slave slave slave do not wire star network follow master slave wiring directions figure 3 26 Master slave slave slave slave do not wire loop network b b good good bad bad Figure 3 12 Installation and...

Страница 29: ...cation 8 on up when IOEM REM device is the last slave on the net work Set dip switch 8 to off down when IOEM REM device is not the last slave on the network Dip Switch configuration table for slave ad...

Страница 30: ...ed one by one as they are added to the network Step 7 Connect Communication Cable for Firmware Update An RS232 communication cable must be connected locally at the expansion module to a PC running Aqu...

Страница 31: ...y get corrupted if any of the following occurs during a firmware update a Firmware Updater window utility is closed b AquaManager is closed c RS232 cable is removed from IOEM REM or PC com port d powe...

Страница 32: ...8 Left click Yes to begin firmware upload Firmware Updater screen opens 9 at window when upload completes 10 Verify uploaded version information is correct at IOEM REM node after manually polling IOEM...

Страница 33: ...e if communication problem is resolved Installation and Wiring Unit is Up To Date indicates that the expansion module has most recent firmware and upload is not necessary 11 Remove RS232 communication...

Страница 34: ...p n 065921 conduit sealer must be used to prevent moisture from entering the IOEM and REM enclosure Apply the sealant after all wiring and connections are complete Failure to use industrial encapsula...

Страница 35: ...IOEM REM Safey and Installation Guide 68 4 Using AquaManager to configure IOEM and REM Channels This page left intentionally blank...

Страница 36: ...ommon terminology used to describe IOEM and REM devices as well as IOEM REM and 5X00 Eaux system functionality I O An IOEM instrument is an I O device whose channels are wired and configured using the...

Страница 37: ...3 modes Each mode is described below 1 Input mode IOEM channels can be configured as an input Input mode selections are digital 0 1Volt 0 5Volt 4 20mAand temperature IOEM inputs use one of the 5X00 in...

Страница 38: ...figuration information see AquaManager Help Contentstab WorkingwithNodes Systemsmanagement Systems Management fields table and link to E Aux bookmarks 7 to upload E Aux system configuration to network...

Страница 39: ...ut channels can only be assigned to 1 5X00 system at a time When selecting a 4 20mA output channel that is currently pending or confirmed AquaManager displays the below 4 20mA Conflict Error window wi...

Страница 40: ...figuration to networked instrument 8 Verify wired 4 20mA output device is regulating the system correctly based on the system value Reset IOEM to factory Follow the steps below to reset an IOEM expans...

Страница 41: ...00 instrument manuals Configuring Sensor Setup Service Output test system menu to verify 4 20mA output Notes At Run Screen 5X00 instrument system using REM relay will show control and or alarm relay i...

Страница 42: ...y AquaManager patch update is sent every 30 days regardless of AquaManager program launch This notification will be visible if after 30 days a new AquaManger version is available regardless of what Aq...

Страница 43: ...t Connector for Auxiliary inputs 3 pole If any parts are damaged or missing contact YSI or your YSI representative im mediately Unpacking and Inspection Appendix 2 Troubleshooting Status leds There ar...

Страница 44: ...ster See System Communica tions Network Master menu 5X00 instrument user manuals more than one device is configured as a master See System Communica tions Network Master menu 5X00 instrument user manu...

Страница 45: ...tion of voltage changes voltage fluctuations and flicker in public low voltage supply systems for equipment with rated current 16A per phase and not subject to conditional connection IEC61010 1 2010 S...

Страница 46: ...r installation If this equipment does cause interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the inter...

Страница 47: ...and proof of purchase transportation prepaid to the Authorized Service Center selected by YSI Repair or replacement will be made and the product returned transportation prepaid Repaired or replaced p...

Страница 48: ...o accept the instrument for service The form may be downloaded from www ysi com by clicking on the Support tab then the Product Return Form Contact Us Appendix 8 Recycling Recycling YSI is committed t...

Страница 49: ...module firmware updater directions 60 F Ferrite bead installation 35 Firmware Updater could not find unit error 65 device timed out error 64 directions to update firmware 60 Unit is Up To Date message...

Страница 50: ...cifications sensors 13 systems non input 13 14 communication 14 RS232 14 firmware 14 hardware relays 13 Stub network wiring 49 Surface mount 5200A 26 Surge and lightning protection 65 Index T Table of...

Страница 51: ...Item 606524 Rev B Drawing 606524 April 22 2015 2015 YSI Incorporated...

Отзывы: