FDO
®
70x IQ H
Maintenance, cleaning, replacement parts
ba76178e01
10/2014
5 - 1
5
Maintenance, cleaning, replacement
parts
5.1
General maintenance instructions
Maintenance condition
We recommend to switch on the maintenance condition each time the
sensor is taken out of the measuring position. This avoids unwanted
reactions of linked outputs. For more detailed information on the main-
tenance condition please refer to the respective IQ
S
ENSOR
N
ET
system
operating manual.
If you would like to disconnect the sensor from the sensor connection
cable, please note the following points:
Before disconnecting the sensor from the SACIQ sensor connection
cable, the sensor has to be cleaned on the outside (see section
5.4.1).
Unscrew the sensor from the SACIQ sensor connection cable.
Place a protective cap on the sensor plug head connector and on
the SACIQ sensor connection cable each time so that no moisture
or dirt can get on the contacting surfaces.
In corrosive environments, close the socket of the sensor connec-
tion cable with the screwable SACIQ-Plug when it is dry in order to
protect the electrical contacts from corrosion. The protective plug is
available as an accessory (see section 5.5 R
).
WARNING
Chemical or biological hazard.
Contact with the sample can be harmful to the user.
Depending on the type of sample, suitable protective
measures must be taken (protective clothing, gog-
gles, etc.).
We do not recommend unscrewing the sensor from the
sensor connection cable in order to do maintenance work.
Otherwise, moisture and/or dirt can get into the plug con-
nection where it can cause contact problems.
Содержание YSI FDO 700 IQ H
Страница 8: ...1 4 ba76178e01 10 2014 Overview FDO 70x IQ H...
Страница 12: ...2 4 ba76178e01 10 2014 Safety FDO 70x IQ H...
Страница 18: ...3 6 ba76178e01 10 2014 Commissioning FDO 70x IQ H...
Страница 26: ...4 8 ba76178e01 10 2014 Measuring Operation FDO 70x IQ H...
Страница 32: ...5 6 ba76178e01 10 2014 Maintenance cleaning replacement parts FDO 70x IQ H...
Страница 40: ...7 6 ba76178e01 10 2014 Technical data FDO 70x IQ H...
Страница 42: ...8 2 ba76178e01 10 2014 Contact Information FDO 70x IQ H...
Страница 47: ......