FDO
®
70x IQ H
Measuring / Operation
ba76178e01
10/2014
4 - 7
4.2.5
Reactivating previous calibration data
The FDO
®
70x IQ H enables you to reactivate the last valid user cali-
bration or the factory calibration. Thus you can immediately go on
measuring if a calibration procedure failed or you suspect that the cali-
bration conditions were not optimally met.
Reactivating
calibration
data
Reactivating old calibration data is only a temporary solu-
tion. Please take into account that the sensor might thus
possibly provide incorrect measured values. Please make
sure the sensor works correctly by carrying out a function
check and/or recalibration.
1
Open the setting table (see section 3.4).
2
In the
Calibration
menu item, select the
User calibration
or
Fac-
tory calibration
setting and then quit the setting table with
Save
and quit
.