OPTICAL SENSOR FOR DISSOLVED OXYGEN FOR CLASS I DIVISION 2 APPLICATIONS

Содержание YSI FDO 700 IQ H

Страница 1: ...OPERATING MANUAL ba76178e01 10 2014 FDO 700 IQ H FDO 701 IQ H OPTICAL SENSOR FOR DISSOLVED OXYGEN FOR CLASS I DIVISION 2 APPLICATIONS...

Страница 2: ...For the most recent version of the manual please visit www ysi com Contact YSI 1725 Brannum Lane Yellow Springs OH 45387 USA Tel 1 937 767 7241 800 765 4974 Email environmental ysi com Internet www y...

Страница 3: ...ommissioning Getting the instrument ready for measuring 3 3 3 4 FDO 70x IQ H setting table 3 4 4 Measuring Operation 4 1 4 1 Measuring 4 1 4 2 Function check and user calibration 4 1 4 2 1 General inf...

Страница 4: ...dous location ratings 7 3 7 5 Electrical data 7 3 7 6 Technical data of the SC FDO 700 7 4 7 7 Technical data of the SC FDO 701 7 5 8 Contact Information 8 1 8 1 Ordering Technical Support 8 1 8 2 Ser...

Страница 5: ...nual has a modular structure like the IQ SENSORNET system itself It consists of a system operating manual and the operating manuals of all the components used Please file this component operating manu...

Страница 6: ...characteristics of the sensor as a whole Thus the sensor can be adapted to the measuring job by exchanging the type of sensor cap The specifica tions of the individual sensor cap types are given in ch...

Страница 7: ...factory calibration Intelligent membrane IQMC technology For each membrane the individual calibration values are determined by a factory calibration and stored on the memory chip of the sensor cap ens...

Страница 8: ...1 4 ba76178e01 10 2014 Overview FDO 70x IQ H...

Страница 9: ...d safety symbols on the product DANGER DANGER indicates an imminently hazardous situation which if not avoided will result in death or serious injury WARNING WARNING indicates a potentially hazardous...

Страница 10: ...he following points for safe operation The product may only be operated according to the authorized use specified above The product may only be supplied with power by the energy sources specified in t...

Страница 11: ...azardous location rated sensor models Hazloc Sensor must be used in hazardous locations Read the name plate on the sensor shaft and verify that the sensor is rated for your specific application DANGER...

Страница 12: ...2 4 ba76178e01 10 2014 Safety FDO 70x IQ H...

Страница 13: ...ensor models Hazloc Sensor must be used in hazardous locations Verify the rating on the product name plate on the sensor shaft The associated apparatus MIQ BB1 YSI part no 207001Y has to be used for c...

Страница 14: ...sensor and the SACIQ sensor connection cable and keep them safe 2 Plug the jack of the SACIQ sensor connection cable onto the plug head connector of the sensor At the same time rotate the socket so th...

Страница 15: ...el designation of the sensor cap SC FDO 70x Ser no Series number of the sensor cap Sensor name The series number of the sensor basis is preset here You can change this entry as necessary by entering a...

Страница 16: ...terized by the Response time t90 This is the time after which 90 of a signal change is displayed The setting range depends on the sensor cap type Calibration Factory calibra tion User calibration acti...

Страница 17: ...ontainer while stirring occasionally until the balanc ing can be performed Salinity On Off Determines whether the entered salinity should be taken into account Salinity input 2 0 70 0 The entry of the...

Страница 18: ...3 6 ba76178e01 10 2014 Commissioning FDO 70x IQ H...

Страница 19: ...measured values appear to be implausible and it is assumed that the service life of the sensor cap is over Routinely within the framework of the company quality assurance Check or calibration medium S...

Страница 20: ...2 cm 1 Switch to the measured value display with M and select the FDO 70x IQ H sensor 2 Press C The next step switches on the maintenance condition for the sensor A corresponding message appears on th...

Страница 21: ...relative slope is outside the tolerance range 0 90 1 10 the check is assessed as erroneous 7 Put the sensor in the measuring position again 8 Wait for the measured value to be largely stable temperatu...

Страница 22: ...e taken into account 1 Switch to the measured value display with M and select the FDO 70x IQ H sensor 2 Press C The next step switches on the maintenance condition for the sensor A corresponding messa...

Страница 23: ...ed before the canceled user calibration Possible results of the user calibration The calibration data is evaluated by the system A calibration can have the following results 1 Open the setting table s...

Страница 24: ...Intens Intensity sufficient too low Res Evaluation of the calibration o k Calibration successful The new calibration data was taken over for measure ment Error Calibration unsuccess ful Invalid calib...

Страница 25: ...ation conditions were not optimally met Reactivating calibration data Reactivating old calibration data is only a temporary solu tion Please take into account that the sensor might thus possibly provi...

Страница 26: ...4 8 ba76178e01 10 2014 Measuring Operation FDO 70x IQ H...

Страница 27: ...crew the sensor from the SACIQ sensor connection cable Place a protective cap on the sensor plug head connector and on the SACIQ sensor connection cable each time so that no moisture or dirt can get o...

Страница 28: ...arly daylight of the interior of the sensor cap will by and by affect the measurement characteristics and shorten the service life of the sensor cap Therefore the interior of the sensor cap should not...

Страница 29: ...brushed off with a soft brush Attention Do not use the brush in the area of the sensor membrane Risk of damage The sensor cap including the sensor membrane should be wiped with a soft and moist micro...

Страница 30: ...emove the sensor cap see section 5 3 2 Clean the sensor head and sensor cap Rinse all inner surfaces with tapwater Remove contamination containing fat and oil with warm water and household washing up...

Страница 31: ...del Order no Cleaning head CH 900 107Y Active valve module does not require a free relay output in the IQ SENSORNET system MIQ CHV PLUS 480 018Y Valve module for the IQ SENSORNET system 182 external r...

Страница 32: ...5 6 ba76178e01 10 2014 Maintenance cleaning replacement parts FDO 70x IQ H...

Страница 33: ...ensor cap as necessary see section 5 3 Dirt inside the sensor cap and sensor head Fixing ring not properly tight ened Sensor cap untight or defec tive Dismantle sensor cap Clean the insides of the sen...

Страница 34: ...bration unsuccessful The sensor is blocked for mea surement As a temporary measure to quickly restore the readiness for service Activate the fac tory calibration see section 4 2 5 For exact measuremen...

Страница 35: ...ure compensation In the range 5 C 50 C 23 122 F 7 2 Application conditions Pressure resistance Sensor with connected SACIQ sensor connection cable The sensor meets all requirements according to articl...

Страница 36: ...there are chloride concentrations of 500 mg L or more Automatic sensor monitoring SensCheck function Monitoring of the membrane function 7 4 Instrument safety 7 4 1 General instrument safety 40 0 400...

Страница 37: ...C22 2 213 The FDO 70x IQ H is a non incendive electrical equipment for use in Class I Division 2 hazardous locations Hazardous location rating 7 5 Electrical data FDO 700 IQ H FDO 701 IQ H Class I Di...

Страница 38: ...t hydrogen sulfide chlorine and ionogenic substances Allowed temperature range Allowed pH range of the test sample 4 12 Calibration procedure Factory calibration by means of IQMC Intelligent Membrane...

Страница 39: ...ine and ionogenic substances Allowed temperature range Allowed pH range of the test sample 4 12 Calibration procedure Factory calibration by means of IQMC Intelligent Membrane Calibra tion procedure U...

Страница 40: ...7 6 ba76178e01 10 2014 Technical data FDO 70x IQ H...

Страница 41: ...g a product for service include the Product Return form with cleaning certification The form must be completely filled out for an YSI Service Center to accept the instrument for service The Product Re...

Страница 42: ...8 2 ba76178e01 10 2014 Contact Information FDO 70x IQ H...

Страница 43: ...5 SC FDO 701 sensor cap SC FDO 701 531 FDO700IQ sensor basis component class adapter ADA Message code Message text EA133x Meas range exceeded or undercut Check process Select other meas range EA233x S...

Страница 44: ...securely Repeat calibration Replace sensor cap EI133x Operational voltage too low Check installation and cable lengths Follow installation instructions Power supply module s overloaded add power suppl...

Страница 45: ...calibration has been replaced by last valid user calibration Caution Wrong measured values possible Carry out a new successful calibration to make sure the sensor operates correctly es correctly IC63...

Страница 46: ...the Einstellungen Settings Service List of all components menu see system operating manual by an automated query from a superordinate process control e g when connected to the Profibus from the IQ Da...

Страница 47: ......