Turb PLUS 2000
Operation
52
100193 Rev 0 ba77254e01 11/2020
NOTICE
The cuvette must be completely dry before it is inserted into the sensor. If there is
any visible moisture present on the cuvette or transducer, there is a great risk of
damaging the sensor electronics and the transducer. Be sure to clean and dry the
cuvette completely just before inserting it into the sensor.
The Vapor Purge system can NOT remove large droplets of water, only residual moisture.
NOTICE
For the Vapor Purge system to function properly, all instrument seals must be
maintained and the desiccant pack must be in good condition (no DESC display).
Figure 10: Operational Parts of the Ultrasonic Cleaning System
Figure 9: Instrument Seals