Setup
, Configuration and Calibration
Turb PLUS 2000
100193 Rev 0 ba77254e01 11/2020
23
Modes of Operation
There are four modes of operation,
AUTO
,
CAL
,
CONFIG
&
HOLD
. To change between AUTO,
CAL, HOLD, and CONFIG use the
and
then the
button to select.
AUTO
mode is the normal automatic operation. This is the default mode when power is applied or
restored. The other three modes of operation are limited to 15 minutes with no key presses after
which they will return to AUTO made operation.
CAL
is the calibration mode. All reading outputs and alarms are held while in this mode of
operation. The instrument was calibrated and tested prior to leaving the factory. Therefore, it is
possible to use the instrument directly out of the box. Under normal conditions, re-calibration is
required once every three months. Quarterly calibration ensures performance within accuracy
specifications.
Relay contacts and the 4-20mA will be frozen (held at the current state) while the instrument is in
the calibration mode
While in the calibration mode, the instrument has a time-out feature that automatically returns the
system operation to the
AUTO
mode after a fifteen (15) minute period of inactivity.
1. Enter the
CAL
mode of the instrument by pressing the
or
buttons until the arrow beside
CAL
is
illuminated, then press the
button.
2. Use the
and
buttons to move forward and backward through the calibration points.
3. To exit the
CAL
mode start a calibration point after the countdown starts press
or
. When the
screen shows ABRT press the
button.
CONFIG
is the configuration mode. In CONFIG mode, you can customize the instrument
according to needs and preferred operation at any time during normal operation. The
CONFIG
mode has been split into sub-menus to facilitate instrument configuration. This is also where
logged files and calibration logs can be downloaded and where new software, when available, can
be uploaded.
All reading outputs and alarms are held while in this mode of operation. Relay contacts and the 4-
20mA will be frozen (held at the current state) while the instrument is in the configuration mode.
While in the configuration mode, the instrument has a time-out feature that automatically returns
the system operation to the
AUTO
mode after a fifteen (15) minute period.
1. Enter the
CONFIG
mode of the instrument by pressing the
or
buttons until the arrow beside
CONFIG
is illuminated, then press the
button.
2. To exit the
CONFIG
mode at any time, press the
button.
3. The
and
buttons are used to move forward and backward through the menus.