Commissioning

Scope of delivery

ORP electrode SensoLyt

®

 900(-P)

Operating manual

Commissioning

Prepare the electrode for measuring as follows:

Remove the watering cap from the electrode tip. Possible salt deposits in 
the area of the watering cap do not affect the measuring characteristics 
and can easily be removed with deionized water.

Connect the electrode to a free IDS sensor plug-in position of the multi 
parameter probe or to an IDS connection of the meter.
To open and close the IDS plug-in position please note the Abschnitt 2.1 
O

PENING

 

AND

 

CLOSING

 

THE

 IDS 

PLUG

 

CONNECTION

Calibrate the electrode according to the operating manual of the meter 
and observe the rules in the following section while doing so:

Preparing the

sensor for

measurement

2.3

Calibration and measurement: General rules

Avoid the carryover of any solution (sample or buffer solution) from one 

Please keep the watering cap. It is required for the electrode to 
be stored. Always keep the watering cap clean.

Connection cables in different lengths to connect the electrode to 
the meter are listed in Abschnitt 7 W

EAR

 

PARTS

 

AND

 

ACCESSORIES

.

SensoLyt

®

 900

Connect the sensor to the meter. 
The sensor is immediately ready to measure.

SensoLyt

®

 900-P

Connect the sensor to a free IDS sensor plug-in 
position of the multi parameter probe or to an IDS 
connection of the meter. To open and close the 
IDS plug-in position please note the Abschnitt 2.1 
O

PENING

 

AND

 

CLOSING

 

THE

 IDS 

PLUG

 

CONNECTION

The sensor is immediately ready to measure.

Connection cables in different lengths to connect 
the SensoLyt

®

 900-P sensor to the meter are listed 

in Kapitel 7 W

EAR

 

PARTS

 

AND

 

ACCESSORIES

.

Connect the sensor to the meter. 
The sensor is immediately ready to measure.

Содержание WTW SensoLyt 900

Страница 1: ...OPERATING MANUAL ba75992e02 02 2017 SensoLyt 900 P IDS pH ELECTRODE WITH POLYMER ELECTROLYTE OPERATING MANUAL...

Страница 2: ...ba75992e02 02 2017 SensoLyt 900 P 2 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...

Страница 3: ...toring the calibration data in the sensor ensures that the correct slope and asymmetry are automatically used if the sensor is operated with different meters On the other hand different calibrated sen...

Страница 4: ...ction Open the locking device step 1 Use your thumb and index finger to press the clips of the connector together and pull the connector out of the plug step 2 Closing the plug connection Make sure th...

Страница 5: ...ring the sensor for measurement 2 3 Calibration and measurement General rules Avoid the carryover of any solution sample or buffer solution from one Please keep the watering cap It is required for the...

Страница 6: ...uring as for calibrating NOTE Prevent contact of the pH membrane to the beaker bottom to avoid scratches on the pH membrane Subsequent calibrations The frequency of subsequent calibrations depends on...

Страница 7: ...e contamination by rinsing with deionized water Other types of contamination have to be removed as follows while the contact time with the detergents should be kept as short as possible NOTE Hydrofluo...

Страница 8: ...nge 0 60 C Membrane resistance at 25 C 500 MOhm Typical application Field Accuracy of the IDS measuring technique Measured parameter Accuracy 1 digit pH 0 004 U mV 0 2 T C 0 1 Pressure range at temper...

Страница 9: ...nnection cable 3 m AS IDS 3 903 851 IDS connection cable 6 m AS IDS 6 903 852 IDS connection cable 10 m AS IDS 10 903 853 IDS connection cable 15 m AS IDS 15 903 854 IDS connection cable 20 m AS IDS 2...

Страница 10: ...Wear parts and accessories SensoLyt 900 P 10 ba75992s02 02 2017...

Страница 11: ......

Страница 12: ...buildings factories and farms In more than 150 countries we have strong long standing relationships with customers who know us for our powerful combination of leading product brands and applications...

Отзывы: