Commissioning, measuring, calibration

SensoLyt

®

 900(-P)

4

ba75992s02

02/2017

2

Commissioning, measuring, calibration

2.1

Opening and closing the IDS plug connection

This section only applies to IDS plug variants (SenTix

®

... -P).

Opening the

 

plug

connection

If necessary, clean the plug connection

Open the locking device (step 1)

Use your thumb and index finger to press the clips of the connector 
together, and pull the connector out of the plug (step 2).

Closing the plug

connection

Make sure that the plug connection is completely dry and clean.

Align the guidance of the connector with the groove in the plug and insert 
the connector in the unlocked plug until it catches (step 3).

Close the locking device (step 4).

2

4

3

1

Plug connection
closed and locked

Guidance

Groove

Содержание WTW SensoLyt 900

Страница 1: ...OPERATING MANUAL ba75992e02 02 2017 SensoLyt 900 P IDS pH ELECTRODE WITH POLYMER ELECTROLYTE OPERATING MANUAL...

Страница 2: ...ba75992e02 02 2017 SensoLyt 900 P 2 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...

Страница 3: ...toring the calibration data in the sensor ensures that the correct slope and asymmetry are automatically used if the sensor is operated with different meters On the other hand different calibrated sen...

Страница 4: ...ction Open the locking device step 1 Use your thumb and index finger to press the clips of the connector together and pull the connector out of the plug step 2 Closing the plug connection Make sure th...

Страница 5: ...ring the sensor for measurement 2 3 Calibration and measurement General rules Avoid the carryover of any solution sample or buffer solution from one Please keep the watering cap It is required for the...

Страница 6: ...uring as for calibrating NOTE Prevent contact of the pH membrane to the beaker bottom to avoid scratches on the pH membrane Subsequent calibrations The frequency of subsequent calibrations depends on...

Страница 7: ...e contamination by rinsing with deionized water Other types of contamination have to be removed as follows while the contact time with the detergents should be kept as short as possible NOTE Hydrofluo...

Страница 8: ...nge 0 60 C Membrane resistance at 25 C 500 MOhm Typical application Field Accuracy of the IDS measuring technique Measured parameter Accuracy 1 digit pH 0 004 U mV 0 2 T C 0 1 Pressure range at temper...

Страница 9: ...nnection cable 3 m AS IDS 3 903 851 IDS connection cable 6 m AS IDS 6 903 852 IDS connection cable 10 m AS IDS 10 903 853 IDS connection cable 15 m AS IDS 15 903 854 IDS connection cable 20 m AS IDS 2...

Страница 10: ...Wear parts and accessories SensoLyt 900 P 10 ba75992s02 02 2017...

Страница 11: ......

Страница 12: ...buildings factories and farms In more than 150 countries we have strong long standing relationships with customers who know us for our powerful combination of leading product brands and applications...

Отзывы: