4.3
BOD determination
Municipal wastewater does not normally contain any toxic or inhibiting sub-
stances. There are enough nutrient salts and suitable microorganisms present.
Under these conditions, BOD determination in the undiluted sample using the
OxiTop
®
(-i/-IDS) measuring system is possible.
Instruments and
accessories
required
OxiTop
®
(-i/-IDS) measuring head
Stirring platform (inductive stirring system)
Thermostat cabinet or box (temperature 20 °C ± 1 °C)
Brown sample bottles (bottle volume 510 ml)
Magnetic stirring rods
Magnetic stirring rod remover
Suitable overflow volumetric flask
Rubber sleeves
Nitrification inhibitor (NTH 600)
Sodium hydroxide tablets (NHP 600)
Further documents
for BOD
determination
WTW application reports (see www.wtw.com)
DIN EN 1899-2
Water quality - Determination of biochemical oxygen demand after n days
(BODn) - Part 2: Method for undiluted samples (ISO 5815:1989, modified);
German version EN 1899-2:1998
4.4
Carrying out the BOD determination
4.4.1 Preparing the measurement
Determining the
sample volume
10.
Exit the setting with
<M>
.
1.
Estimate the expected BOD value for the waste water sample.
The expected BOD value is approximately 50% of the COD value.
2.
Look for the relevant measuring range in the following table and take
the values for sample volume and multiplier.