Why calibrate?
During the operation of a pH sensor, the slope and asymmetry of the sensor
change with time. The calibration procedure determines the current slope and
asymmetry of the sensor.
When to calibrate?
Calibrate before measuring and at regular intervals (depending on the
application).
Calibration
procedures
The
CAL TEC AUTO
calibration procedure enables a fully automatic calibration
with Technical buffer solutions.
Technical buffer solutions, see chapter 4.5
.
The
CAL CON 2P
calibration procedure enables the conventional 2-point
calibration with two different buffers.
The calibration procedure
CAL CON 1P
enables the conventional single-point
calibration with any one buffer.
4.2.2 Calibration procedure on the IQ S
ENSOR
N
ET
Before starting, make sure that the correct calibration procedure is
Prepare the calibration standards.
Step 1:
The sensor is in the sample.
The procedure is started from the measured value display with
<C>
(see IQ S
ENSOR
N
ET
system operating manual).
During calibration, the sensor is in maintenance condition.
All linked outputs remain in their current status.
When operating the sensor in a flow through armature:
Remove the sensor from the flow armature.
Step 2:
Calibration:
Follow the instructions on the display.
For details on the selected calibration procedure, see:
Once the calibration routine is complete, the display returns to the
measured value display.
The measured value flashes. The sensor is still in maintenance
condition.