Messages
This chapter contains a list of all the message codes and related message texts
for pH sensors.
All message codes of IDS pH sensors end with the number "313".
4.4.1 Error messages
4.4.2 Informative messages
4.5
Accessories
IDS sensors
For general information on
the log book and message codes, see
IQ S
ENSOR
N
ET
system operating manual.
Message code
Message text
EC1313
Sensor could not be calibrated,
Sensor blocked for measurement
* Check calibration conditions and calibration standard
* View calibration history
* Service sensor immediately
(see operating manual)
Message code
Message text
IC1313
Sensor has been successfully calibrated
* For calibration data, see calibration history
IC4313
Last valid calibration has been activated. Make sure the
sensor operates correctly.
Description
Model
Order no.
IDS pH sensor
Sentix
®
945
103743
Sentix
®
945-P
103764
IDS connecting cable, 1.5 m
AS/IDS-1,5
903850
Adapter for the connection of sensors
with S7 plugs
ADA S7/IDS
108130