Commissioning
FCML 412 N(-S), FCML 412-M12
10
ba77197e02
01/2024
3.3
Electrical connection
How to connect the FCML 412 N(-S) sensor to the measuring transmitter or
the MIQ/IDS module is described in the operating manual of the measuring
transmitter.
How to connect the sensor FCML 412-M12(-x) to the MIQ/IDS module is
explained in the operating manual of the module MIQ/IDS.
3.4
Installation in the flow through vessel D-CL
The flow-through vessel
D-CL
is optimized for the FCML 412 N(-S) chlorine
sensors.
Proceed as follows to install the sensor:
The position in the flow through vessel has a considerable effect
on the approach flow and, hence, the measuring function of the
sensor. Therefore carefully position the sensor as described
below.
1.
Insert the black O-ring, then the PVC slide ring in the 1" installation
opening of the flow through vessel.
2.
Push the coupling ring onto the sensor.
3.
Insert the sensor with the coupling ring in the flow-through vessel.
4.
Screw the sensor tight with the coupling ring.