22/24
9.4.5 Calibration error
In each device predefined ranges for the calibration values are stored. If the determined calibration value is out of
the range the message
!Out of limit!
appears in the calibration information CAL INFORMATION. The deter-
mined calibration value will be stored nevertheless, so that the measuring and controlling is secured. As the case
a new calibration or a replace of the sensor is necessary.
10 Specifications
Configuration
directly with the device by 5 Keys and Display (plain text menu structure) or by
means of the PC interface and corresponding configuration program
Display
graphic LCD Display, 128 x 64 Pixel, backlight
Interface
USB with
„Interface HMG USB”, isolated
Power supply
3 Batteries AA 1,5 V
Ambient temperature
-10...55 °C
Relative humidity
max. <95 % (not condensing)
Housing
protection IP 65, dimensions 200 x 90 x 40 mm (L x W x H)
Weight
approx. 260 g (with battery)
EMC
acc. EN 61326 class B
ESD (EN 61000-4-2)
Air:
±8 kV
Contact:
±4 kV
Criterion B
Radiated Radio frequency (EN 61000-4-3)
Test level:
10 V/m
Citerion A
Burst (EN 61000-4-4)
I/O Signal:
±1 kV; 5/50 ns; 5 kHz
Criterion B
Emissions standards (EN 55011)
Radiated and conducted emission
Class B
Connectors
connector BK for Sensor
4-pole USB Interface socket
Housing
ABS, IP 65
Dimensions
200 x 95 x 40 mm (B x H x T)
Weight
290 g incl. batteries
Measuring range TM 40
pH -2.00...16.00, -1999...1999 mV
Temperature: -10.0...100.0 °C
Measuring range AM 40
0...200.0 %, 0...20.00 mg/l
Temperature: -10.0...100.0 °C
Measuring range LF 40
0...200.0 µS/cm
, TDS 0 …200 mg/l
0...2000 µS/cm, TDS 0...2000 mg/l
0...20.00 mS/cm, TDS 0...20 g/l
0...500.00 mS/cm, TDS 0...500 g/l
automatic range
calculated resistance value
calculated salinity value
Temperature: -10.0...100.0 °C
Recommended electrodes/sensors suitable for connection to the controller in accordance with the individual de-
tailed technical data sheets.