Select the position you would like to use by using the buttons
or
. Press the
-button to accept
and save the entry. To cancel the input dialogue without saving the value press
+
.
8.4 Calibration oxygen
Menu structure
Passwort
pass word
Kalibrieren
calibrate
Temp.Offs.
temp. offs.
Einpunkt
one point
Automatik
automatic
zurueck
back
Anstieg
slope
B>20°C
zurueck
back
Eingabe
input
B<20°C
CAL INFORMATION
Open with
+ password the calibration menu.
All kinds of calibration method will be displayed. Choose now your method.
Please take note for all calibration method the temperature compensation.
8.4.1 Data input
Data input stands for the input of the specific values of the connected sensor, for example as determined at the
laboratory before.
Select the corresponding calibration value whose settings you want to change. An input dialogue will now enable
you to change the values within corresponding limits. If you exceed these you will get an error message to
prompt you to enter a value within the fixed limits.
Select the position you would like to change by using the buttons
or
and change with
or
.
Press the
-button to accept and save the entry.
To cancel the input dialogue without saving the value press
+
.
8.4.2 Settings of the Temperature Offset
This menu item enables you to set a temperature value offset. For this purpose open the menu item
„temp.offs”.
Now, an input dialogue will be displayed which facilitates the setting. The offset can be positive or negative.
Select the position you would like to change by using the buttons
or
and change with
or
.
Press the
-button to accept and save the entry.
To cancel the input dialogue without saving the value press
+
.
back