157
Command
Description
Reply
aaAA
automatic allocation of device address
aaY
aaMC1...XX
choosing a method
aaY
aaBF
„filling burette“. Aufsatz wird gefüllt.
aaY
aaBV
output of dosed volume in ml
aa0.200
aaDA
dose volume without filling, with adding the volume
aaY
aaDB
dose volume without filling, reset of the volume
aaY
aaDO
dose volume with filling, without adding the volume
aaY
aaGDM
dosing speed in ml/min
aaY
aaGF
filling time in seconds (min is 20, default 30)
aaY
aaES
„ESC" function one step backwards
aaY
aaEX
„EXIT“ function.back to main menu
aaY
aaGDM
dosing speed in ml/min (0.01
– 100 ml/min)
aaY
aaGF
filling time in sec (adjustable 20
– 999 seconds)
aaY
aaGS
output serial no. Of device
aaGS08154711
aaLR
output report (short report)
aaY
aaLI
output method content
aaLO
output documentation (as configured)
aaRH
request of identification
aaIdent: TITRONIC
®
500
aaRC
send last command
aa“last command“
aaRS
report status
aaStatus:
“text
possible answers are:
„STATUS:READY“ for ready
„STATUS:dosing“ dosing
„STATUS:filling“ filling
„ERROR:busy“ if no interchangeable unit has been attached
aaSM
start selected method
aaY
aaSEEPROM
EEPROM reset to factory defaults
aaY
aaSR
stop the actual function
aaY
aaSYS5
adjust language to „German“
aaY
aaSYS1
adjust language to „English“
aaY
aaSYS2
adjust language to „French”
aaY
aaSYS3
adjust language to „Spanish”
aaY
aaVE
Version number of the software
aaVersion: