Maintenance /Cleaning of the electrode
31
Cleaning of the electrode
Use clean drinking water for cleaning the electrode.
A 10% hydrochloric acid solution can be used for cleaning if the electrodes are
greasy. Protein coatings are removed with Pepsin cleaning fluid.
Electrode lifetime
The expected lifetime of the electrodes is dependent of the measurement. The
lifetime is given with reservation as the physical conditions as well as the
temperature, pressure and pollution of the fluid has great influence on the
stated lifetime.
The lifetime is based on
77° F / 25° C
, but if the temperature is doubled, the
lifetime is halved in worst case!
pH electrodes in purifying plants for ordinary household waste-water lasts
approximately one year, whereas the lifetime of pH electrodes in industrial
plants is dependent on measuring application and the process. Redox
electrodes have a lifetime of approximately two years, depending on cleaning
intervals.
Electrode spare parts
Item no. Description
160310
pHix® pH electrode, 0-point=4,6 pH
160311
pHix® pH electrode, 0-point=7,0 pH
160312
pHix® redox electrode
571030
Black rubber gasket for electrode
521409
Union ring w. Viton gasket for mounting of pHix® on pipe
- NOT
AVAILABLE IN THE US
521441
Viton gasket for union ring
- NOT AVAILABLE IN THE US
The following is recommended for cleaning and buffer adjustment:
Содержание MJK pHix Compact V3
Страница 1: ...Manual EN 4 10 Version 2103 pHix Compact V3 NG NEXT GENERATION...
Страница 2: ...pHix Compact V3 NG next generation EN 4 10 Version 2103 2 Your notes...
Страница 15: ...pHix Compact V3 NG next generation EN 4 10 Version 2103 Mounting Mechanical mounting 15...
Страница 22: ...pHix Compact V3 NG next generation EN 4 10 Version 2103 Mounting Connection examples 22 Connection to PLC_PLS...
Страница 87: ...This page has been intentionally left blank...