Turb 430 IR/T
Technical data
51
ba75507d09
08/2021
7.2
Turbidity
7.2.1
Turb 430 IR
7.2.2
Turb 430 T
Measuring principle
Nephelometric measurement according to DIN EN ISO 7027
Light source
Infrared LED
Measuring range
0.01 ... 1100 FNU/NTU
Resolution
In the range
0.01 ... 9.99
max. 0.01 FNU/NTU
In the range
10.0 ... 99.9
max. 0.1 FNU/NTU
In the range
100 ... 1100
max. 1 FNU/NTU
Accuracy
In the range
0 ... 1100 FNU/NTU
± 2 % of the measured value
or ± 0.01 FNU/NTU
Repeatability
0.5% of the measured
value
Response time
4 seconds
Calibration
Automatic 3-point calibration
Measuring principle
Nephelometric measurement according to US EPA 180.1
Light source
White light tungsten lamp
Measuring range
0.01 ... 1100 NTU
Resolution
In the range
0.01 ... 9.99
max. 0.01 NTU
In the range
10.0 ... 99.9
max. 0.1 NTU
In the range
100 ... 1100
max. 1 NTU
Accuracy
In the range
0 ... 500 NTU
± 2 % of the measured value
or ± 0.01 NTU
In the range
500 ... 1100 NTU
± 3% of measured value
Repeatability
1% of the measured
value
Response time
7 seconds
Calibration
Automatic 3-point calibration