Operation
Turb 430 IR/T
30
ba75507d09
08/2021
4.7.1
Settings for calibration
Settings
The settings are in the menu
Konfiguration
/
Trübung
/
Kalibrierung
.
To switch to the
Konfiguration
menu, press the
<MENU>
key.
FreeCAL
2 ... 5
Guided flexible calibration with 2 to 5
freely selectable calibration standards.
The nominal values for the calibration
standards are entered manually.
The order of the entered nominal values
corresponds to the order of the calibra-
tion standards during the calibration
process.
Calibration
type
Standards
(FNU/NTU)
Explanation
Menu item
Possible
setting
Explanation
Kalibrierprotokoll
-
Display the calibration record of
the last calibration.
Kal.-typ
3-P StdCAL
FreeCAL
QuickCAL
The selected calibration type is
used to calculate the current
measured value.
The active calibration type is indi-
cated in the measured value dis-
play, and stored together with a
measured value.
When you start calibration with
the
<CAL/ZERO>
key, it is al-
ways started with the selected
calibration type.
Standard
(only with
Kal.-typ
FreeCAL
)
Anzahl
1 ... 5
Number of calibration standards
for the
Kal.-typ
FreeCAL
Standard
1
...
Standard
5
Here the nominal values are en-
tered for all calibration standards