background image

 

 

26   

 

 

 

°C

 -default- 

 

°F 

 

P9.2 Date and time setting 

Access this setup menu to update the device date and time. 

Use the directional keys to change the year, confirm with 

 and repeat the same operation for month, 

day, hours and minutes. 

P9.3 Backlight Mode 

Access this setup menu to select the contrast mode to use for the display backlight: 

 

INDOOR (In)

 – Recommended if you use the device indoors. 

 

OUTDOOR (Out)

 – Recommended if you use the device outdoors. 

 

AUTOMATIC (Auto)

 – Default option. Thanks to the brightness sensor, the display automatically 

adapts to the environment conditions. This mode also ensures longer battery life. 

P9.4 Luminosità

 

Accedere a questo menu di setup per scegliere tra tre differenti livelli di luminosità del display: 

 

 

LOW

 – bassa 

 

NORMAL

 – media 

 

HIGH

 – alta 

 

Note

: Keeping the display bright always adversely affects battery life 

P9.5 Sleep mode 

Access this setup menu to select whether and after how long activating the device Sleep mode: 

 

 

OFF

: Sleep mode off. 

 

2 MIN

: The instrument enters Sleep mode if no key is pressed for 2 minutes. 

5 MIN

: The instrument enters Sleep mode if no key is pressed for 5 minutes. 

 

When the device is in Sleep mode, the brightness of the display is reduced to a minimum, significantly saving 

battery consumption. 

 

Note

: Sleep mode only affects the brightness of the display. All other instrumental functions continue to 

operate normally (e.g. Data Logger). 

To exit from the Sleep mode and return the display to normal brightness, press ANY button. 

Once the display brightness is activated, the buttons reacquire their function (paragraph "Key function"). 

 

P9.8 Reset Settings  

Access this setup menu to restore the instrument to factory conditions. 

 

IMPORTANT:

 Restoring the factory parameters does not delete the stored data. 

 

 

P9.9 Auto off 

Access this setup menu to activate or deactivate the auto-shutdown of the instrument: 

 

YES: 

The instrument automatically turns off after 

20 minutes 

of inactivity.

 

 

NO:

 The instrument remains always on, even if you are not using it.

 

 

Содержание OXY 70

Страница 1: ...1 Rev 1 0 26 02 2020 EN OXY 70 Vio INSTRUCTIONS MANUAL O2 mg l mbar Temp...

Страница 2: ...2 Rev 1 0 26 02 2020 EN...

Страница 3: ...eyboard 10 LED 10 Installation 10 Supplied components 10 Start up 11 Connection of the power supply 11 Power on date and time update power off 11 Replacement of batteries 12 Instrument transportation...

Страница 4: ...ophore 22 Data Logger Function 23 Setup for Data Logger parameter 23 Composition of setup menu for the Data Logger menu 23 Example of automatic Data Logger mode 24 Example of manual Data Logger mode 2...

Страница 5: ...olour LCD display shows all the necessary information such as the measurement the temperature the buffers used for the last calibration also custom the condition of stability Everyone can use these to...

Страница 6: ...zardous situation which could lead to serious injury or death if not avoided ATTENTION for a dangerous situation with reduced risk which can cause material damage data loss or minor or medium sized ac...

Страница 7: ...ular functionality of the device and the operator safety are guaranteed only if all the following indications are respected the instrument can be used in accordance with the specifications mentioned a...

Страница 8: ...nd uses the tool or authorizes its use by other people is the owner of the tool and is responsible for the safety of all users of the tool and third parties The owner of the instrument must inform use...

Страница 9: ...Sound level during standard operation 80 dB Environmental operating conditions 0 60 C Maximum permissible humidity 95 non condensing Maximum altitude of use 2000 m System dimensions 185 x 85 x 45 mm...

Страница 10: ...ts The instrument is always supplied inside the specific carrying case with these accessories batteries adaptor 5V with USB cable standard 0 oxygen paper tissues screwdriver beacker multilingual user...

Страница 11: ...by pressing the button The display initially activates all segments and then appears model and firmware of the instrument settings relating to the most important parameters and possible information a...

Страница 12: ...original case only to transport the instrument If you need to buy it again contact your local distributor The interior of the case is shaped to be able to house the instrument and the sensors still c...

Страница 13: ...he keys regain their function Inputs Outputs connections Use original accessories guaranteed by the manufacturer only If necessary contact your local distributor OXY 70 Vio top panel READ THE MANUAL B...

Страница 14: ...n before turning off To scroll through the different parameter screens press the key Sequence of parameters in measure mode OXY 70 Vio O2 mg l mbar Note Pressing the button after the last parameter th...

Страница 15: ...etup menu In measure mode press the key to enter SETUP mode select the parameter to edit by using the directional keys and confirming with OXY 70 Vio DO SETTINGS LOG SETTINGS SETTINGS Within the selec...

Страница 16: ...Mode P9 8 Reset Settings P9 9 Auto Power Off Temperature measurement ATC MTC ATC The direct measurement of the sample temperature for all parameters is carried out through the NTC 30K probe integrate...

Страница 17: ...TA P5 7 SET DUE CAL NO HOURS DAYS NO P5 8 RESET SETTINGS YES NO NO P5 9 TEMPERATURE CAL YES NO P5 1 Cal 0 Calibration with Standard Zero 02 Access this setup menu to select the calibration with Standa...

Страница 18: ...nge the number that appears in the center of the display entering the hours or days that must elapse between two settings and confirm with If a calibration deadline is set the icon is present on the d...

Страница 19: ...ght of blue colour that is reflected from the luminophore on an inner photocell When the oxygen permeating through the membrane comes in contact to the luminophore it modifies the blue light in propor...

Страница 20: ...ore is replaced by a new one The probe is not used for a long period of time 3 6 months The instrument does not calibrate at 100 in this case calibrate it before at 0 The instrument does not measure c...

Страница 21: ...sed for a long time After the luminophore replacement Errors during calibration NOT STABLE The button was pressed with still unstable signal Wait for the icon to appear to confirm the point WRONG BUFF...

Страница 22: ...eading does not stabilize it is necessary to perform a maintenance of the probe in the following way 1 Check that the luminophore is clean in case wash it with water 2 The luminophore must be intact u...

Страница 23: ...o the USB port on the top panel of the instrument and the other end to a COM port on the computer Use the USB cable supplied with the instrument only Setup for Data Logger parameter In measure mode pr...

Страница 24: ...to MINUTES Use the directional keys to change the number that flashes on the display Enter 2 and confirm with Go back to measure mode The icon is lit in the lower string of the display which indicates...

Страница 25: ...t delete the stored data Instrument Setup menu In measure mode press key to access the SETUP menu Use the directional keys to move to SETTINGS P9 0 and access the menu by pressing the key Move with th...

Страница 26: ...after how long activating the device Sleep mode OFF Sleep mode off 2 MIN The instrument enters Sleep mode if no key is pressed for 2 minutes 5 MIN The instrument enters Sleep mode if no key is presse...

Страница 27: ...switch on the instrument Wait for connection to be established the connection data are shown at the bottom left of the display Functions Download the data saved in the instrumental memory are downloa...

Страница 28: ...ll repair or replace defective components This warranty does not apply if the product has been damaged used incorrectly exposed to radiation or corrosive substances if foreign materials have penetrate...

Отзывы: