XO CARE A/S

YB-223 Version 3.00

XO Osseo, User Guide

9

Mounting the irrigation tubing system

1. The non-sterile assistant removes the magnetically retained pump cover in order to expose the

pump rotor and the bracket with two (upper and lower) U-shaped cut-outs.

2. The non-sterile assistant opens the tubing package, and the sterile assistant takes out the main

tubing.

3. While held in position by the sterile assistant, the non-sterile assistant stretches the pump part

around the pump rotor and fits the two plastic connectors into the respective U-shaped cut-

outs.

See Illustration below.

Fig. 2

CAUTION:

The pump tubing part must be correctly and

carefully stretched around the pump rotor with that plastic connector which

connects with the inlet part, placed in the upper U-cut-out.

4. The non-sterile assistant opens the separate sub-package including the two outlet tubings, and

the sterile assistant selects the desired outlet tubing necessary for the operation concerned (single

or combined burr cooling).

5. The  sterile  assistant  removes  the  protective  end  cap  on  the  male  Luer  lock  connector  on  the

transport part and connects the desired outlet tubing therewith.

6. The sterile assistant connects the free end or ends of the outlet tubing with respective irrigation

inlets on the contra-angle for internal, external or combined burr cooling as the case may be.

7. The  non-sterile  assistant  removes  the protective  end  cap  from  the  bag cannula,  connects  the

cannula with a sealed bag or bottle with sterile irrigant, and suspends the bag or bottle in the bag

hanger.

The irrigation system is now ready for operation.

Содержание OSSEO

Страница 1: ...XO OSSEO User Guide YB 223 Version 3 00...

Страница 2: ...1 1 STARTING PROCEDURE 10 D 1 2 PROGRAMME STEP PRELIMINARY DRILLING 11 D 1 3 PROGRAMME STEP FINAL DRILLING 11 D 1 4 PROGRAMME STEP TAPPING FORWARD SCREW IN 12 D 1 5 PROGRAMME STEP TAPPING REVERSE SCR...

Страница 3: ...implantology in particular In your own interest we kindly ask and recommend that you take the time to carefully study this manual in order to familiarize yourself with the XO OSSEO SYSTEM and its ope...

Страница 4: ...TION AND COMPONENTS XO OSSEO SYSTEM 1 Control box 2 Pump Cover 3 Handmotor with cord Contra angle 20 1 must be ordered separately 4 Bag hanger 5 System keys 6 Adjustment keys 7 Foot control 8 Tubing k...

Страница 5: ...s for tubing cord hook up Prior to use please check that these parts are present and are free of any damage A 2 SUPPLEMENTARY ACCESSORIES The following extras are available but are to be ordered separ...

Страница 6: ...ction gearing and accordingly the respective torque and speed values on the display will only be true when that particular reduction gearing is used For implantology we recommend the Micro M ga 20 IMK...

Страница 7: ...ll socket with appropriate protective ground connection 2 Connect the foot control by plugging its cable connector end into the socket on the rear of the control box Fig 1 3 Insert the bag hanger into...

Страница 8: ...THE IRRIGATION TUBING SYSTEM The XO OSSEO SYSTEM is delivered with irrigation tubing kits in pre sterilized packages The tubings are for one time use only Do not re sterilize The tubing kit is design...

Страница 9: ...connector which connects with the inlet part placed in the upper U cut out 4 The non sterile assistant opens the separate sub package including the two outlet tubings and the sterile assistant selects...

Страница 10: ...PLANT SYSTEM PARAMETERS The following instructions only relate to operation with system specific parameters as factory programmed in accordance with respective implant manufacturer s recommendations I...

Страница 11: ...ON OFF switch on the foot control see Fig 4 NOTE The irrigation pump starts about 1 second before the handmotor in order to ensure ample irrigation The pump and handmotor will stop as soon as the foot...

Страница 12: ...ing for and screwing in of threaded implants D 1 5 PROGRAMME STEP TAPPING REVERSE SCREW OUT Enter step by stepping once again on the select switch on the foot control As shown in Fig 8 the parameters...

Страница 13: ...factory programmed parameters on the display i e speed 25 RPM direction forward and torque 25 Ncm Irrigation is switched OFF 2 Start the handmotor by stepping the motor start switch on the foot contr...

Страница 14: ...stored and memorized by clicking the central adjustment port until the cursor vanishes from the display NOTE The handmotor cannot start until altered parameters have been stored in this way cursor van...

Страница 15: ...A 2 of this manual The parameters speed RPM torque Torque and irrigation flow rate H2O can be adjusted individually NOTE The torque setting is displayed in percent of the maximum torque output for the...

Страница 16: ...ontrol 1 Enter the adjustment mode by stepping once on the select switch on the foot control A cursor white background on the display and a plus sign now indicate that the speed is open for increasing...

Страница 17: ...her by clicking the central port until the cursor vanishes or by stepping once on the motor ON OFF switch on the foot control You are now ready to operate with your own parameter settings Start the ha...

Страница 18: ...e may cause damages to the handmotor CAUTION Do not lubricate through the open front end of the handmotor Double wrapping is recommended Sterilize only by steam autoclaving at a max temperature of 135...

Страница 19: ...2 speed range 600 80000 rpm 260220 Surgery system access key 1 1 speed range 300 40000 rpm 260222 Surgery system access key 2 1 speed range 150 20000 rpm 260224 Surgery system access key 10 1 speed ra...

Страница 20: ...f EN 60601 1 IEC 601 1 and EN 60601 1 2 class B Technical changes including repairs adjustments or calibrations and modifications may only be made by XO CARE A S or specialists approved by XO CARE A S...

Страница 21: ...ed in the chuck of the handpiece head The handmotor does not operate 1 Make sure that the motor cord connector is pushed fully and firmly into its socket on the front of the control box 2 Make sure th...

Страница 22: ...ubing is not sufficiently stretched around the pump rotor CAUTION The items above represent the problems that can be overcome without sending the unit for repair Do not attempt to repair or service th...

Страница 23: ...nt Mains voltage 115 230V 50 60 Hz Mains connection 1 phase with protective ground Max power consumption 380 VA Fuses 2 pcs T 3 15A 115V 2 pcs T 1 6AL 230VAC Replacement of fuse Fuse carrier is remove...

Страница 24: ...will only be carried out upon receipt of the defect unit together with its Guarantee Card Responsibility as to the safety reliability and performance of the XO OSSEO SYSTEM will not be accepted by XO...