XO
®
4 User Guide
YB-253 Version 3.10
16
xxx
Figure 31 Replacement of bulb in micro motor
Please refer to the user guide for the micro motor.
5.5
Ultrasonic scaler
Please refer to manufacturer’s documentation.
5.6
Light curing lamp
Please refer to manufacturer’s documentation.
5.7
XO 4 Lamp, halogen type
Replace the bulb (ref.no. WH-007) as follows:
1 Switch off XO 4 unit.
2 Remove the splash screen by means of the 2 screws.
3 Push and turn the shade counter-clockwise and remove it.
4 Avoid touching the bulb with your fingers. Use clean gloves
or a clean piece of cloth when handling the bulb.
5 Use the plastic tool to carefully mount the bulb in the socket.
6 Mount the shade and the splash screen again.
7 Switch on XO 4 and activate XO 4 Lamp.
xxx
Figure 32 Replacement of lamp bulb
5.8
XO 4 Foot Control
If XO Foot Control slides on the floor when activated: Clean rub-
ber feet under foot control with benzine.
Figure 33 Cleaning of rubber feet
Clean rubber feet under XO Foot Control regularly
with benzine.
5.9
Hand instrument table, adjust-
ment
The inclination of hand instrument table can be adjusted with a
4 mm Allen key.The inclination of the hand instrument table can
be adjusted by means of a 4 mm Allen key:
Figure 34 Adjustment of hand instrument table
1