Do not wash unit instrument hoses in a thermo dis-
infector!
In the event that the instrument hoses have been
removed, one must ensure that the plugs and sockets are
completely dry before the hoses are re-mounted on the in-
strument bridge again!
4.5.3
Multifunctional syringe
The syringe cover and the tip can be autoclaved at max. 134°C.
Press the button at the back (1) and slide off the cover. Remove
the tip by loosening the joint (2).
xxx
Figure 25 Removal of syringe cover
Do not wash the multifunctional syringe in a thermo
disinfector and do not autoclave!
4.5.4
Bien Air MC3 micro motor
The micro motor cover (1) can easily be removed for autocla-
ving at max. 134°.
Do not pull the tube!
Disinfect the micro motor with Dürr FD-322.
Lubricate the o-rings (2) with XO silicone grease (ref. no. YR-
002)
xxx
Figure 26 Removal of micro motor cover
Do not wash the micro motor in a thermo disinfector and do
not autoclave!
4.5.5
Other instruments
For turbines, ultrasonic scalers, LC lamps, intraoral video came-
ras and other instruments:
Please refer to the user manual supplied with the instrument.