microDXP Technical Reference Manual
Version 3.15
October 7, 2019
11
Multi-channel analysis with up to 8K (8192) bins, allowing for optimal use of data to
separate fluorescence signal from backgrounds.
Instantaneous throughput beyond 1,000,000 counts per second (cps) into the
spectrum for the standard 40MHz ADC clock speed.
Digital trapezoidal filtering, with programmable peaking times between 100 ns and
24
s for the standard 40MHz ADC clock speed.
Digitally-controlled analog gain: 25.5 dB range, 16 discrete settings.
Digital fine gain trim: 16-bit precision
Pileup inspection criteria computer settable, including fast channel peaking time,
threshold, and rejection criterion.
Accurate ICR and live-time reporting for precise dead-time corrections.
1.1.2 Embedded Systems Features
Data acquisition and control via RS-232 or USB.
Twenty-four (24) sets of optimized spectrometer parameters, or PARSETs, are stored
and retrieved on a per-peaking-time basis in nonvolatile memory.
Five (5) sets of MCA parameters, or GENSETs, are stored and retrieved in
nonvolatile memory.
XIA provides the ProSpect software application for microDXP parameter set
configuration—intended for use both in evaluation and production phases.
1.1.3 Custom OEM Features
Customized firmware development for special applications such as time-resolved
spectroscopy, multiple SCA’s and pulse-shape analysis.
Flexible auxiliary digital I/O: 4 general purpose lines, a Gate signal to externally
control data acquisition, I2C bus interface and an external interrupt line.
Assembly options are offered to exclude various hardware and software features in
order to reduce the cost for dedicated applications (further details in the following
section).
1.2 Hardware Requirements
1.2.1 Host Computer
The microDXP can communicate with any host computer/controller via RS-232. Windows
7 or later is required to operate via USB.
1.2.2 Detector/Preamplifier
The microDXP accommodates nearly all detector preamplifier signals. The two primary
capacitor-discharge topologies, pulsed-reset and resistive-feedback, are supported. The
voltage compliance range in the DXP analog circuitry imposes the following constraints:
Parameter
Minimum
Maximum*
X-ray pulse-height
50
V
125 mV
Input voltage range
-
+/- 4.0 V
Decay time
100 ns
8.19 ms
Table 1-1: Analog input signal constraints