XFT Medical G4 Скачать руководство пользователя страница 15

 

Gait mode interface 

 

Start gait mode 

 

– 

Under the gait mode, user can choose among smart mode, normal mode, and manual mode. Click the “smart mode” 

position to pop up the mode selection list and select one. 

a. Smart mode: The Stim Unit automatically calculates the tilt angle A of the electrical stimulation and the tilt angle 

B of shutting down electrical  stimulation according to the gait data of the first four steps of the patient. The 

parameters that can be adjusted are the electrical stimulation intensity, frequency and pulse width in the lower level 

interface of the “Parameter Settings”. 

b. Normal mode: The Stim Unit performs electrical stimulation according to the set parameters. The parameters that 

can be adjusted are the electrical stimulation intensity, and the frequency, pulse width, tilt angle A, tilt angle B, 

duration, delay time, ramp up and ramp down in the lower-level interface of the “Parameter Setting”.   

c. Manual mode: After starting this mode, the Stim Unit performs electrical stimulation according to the electrical 

stimulation button of the interface; press the button to start electrical stimulation, stop the electrical stimulation by 

the release, and continue to hold the electrical stimulation for continuous output. The parameters that can be 

adjusted are the electrical stimulation intensity and the frequency and pulse width in the lower level interface of the 

“Parameter Settings”. 

 

Training mode 

– 

The training mode performs electrical stimulation based on a combination of parameters of the selected training 

mode. The parameters combine 9 fixed combinations and one custom. The electrical stimulation can be adjusted 

after the fixed mode starts, and other parameters cannot be adjusted. All parameters can be adjusted under custom. 

– 

Start: The App sends the training mode parameters and mode start command to the Stim Unit, receives the Stim 

Unit reply, and enters the training work interface. 

– 

Pause: The App sends a training mode pause (stop) command to the Stim Unit, receives a Stim Unit reply, and the 

training is suspended. 

– 

Continue: The App sends the training mode (start), at which point the current intensity given in the parameter 

settings is the current strength before the pause. After receiving the Stim Unit response, the training work continues. 

– 

Stop: After long press and stop for 2s, send the training mode stop command to the Stim Unit, receive the Stim Unit 

reply, and the training mode stops. 

Содержание G4

Страница 1: ...Foot Drop System User Manual Caution Federal law restricts this device to sale by or on the order of a practitioner licensed by the law of the State in which he she practices to use or order the use...

Страница 2: ...and Safety 3 2 Overviews 8 3 Product Illustration 9 4 Operation Instruction 12 5 Attentions 17 6 Care and Maintenance 17 7 FAQ Troubleshooting 18 8 Product Configuration 18 9 Product classification 19...

Страница 3: ...the device battery accessory packing in accordance with the legal obligation in your area Contraindications Powered muscle stimulators should not be used on patients with cardiac demand pacemakers Do...

Страница 4: ...ds 10mA cm2 r m s Please use with caution if the used areas have structural deformity This product should be conducted by doctors Patients should keep stable and not move while using this unit Patient...

Страница 5: ...ctions may put the user at undue risk of injury Adverse Reactions Skin irritation and burns beneath the electrodes have been reported with the use of powered muscle stimulators Electromagnetic Compati...

Страница 6: ...is used in such an environment Immunity test IEC60601 test level Compliance Guidance Electrostatic discharge ESD IEC 61000 4 2 6kV contact 8kV air 6kV contact 8kV air Floors should be wood concrete or...

Страница 7: ...ermined by an electromagnetic site survey a should be less than the compliance level in each frequency range b interference may occur in the vicinity of equipment marked with the following symbol Radi...

Страница 8: ...s the lack of ankle dorsiflexion in patients who have sustained damage to upper motor neurons or pathways to the spinal cord During the swing phase of walking the XFT 2001EA electrically stimulates th...

Страница 9: ...for 1 4 hours a day every morning and evening 20 minutes each time 3rd week later Walk for 4 8 hours a day every morning and evening 20 minutes each time Note take off the cuff for 15 minutes after e...

Страница 10: ...working states of the unit such as gait mode training mode electrode loose low battery icon electrical stimulation output icon electrical stimulation intensity etc 3 2 2 Indicators Power on indicatio...

Страница 11: ...is low there will be a battery icon display battery display zero capacity at low voltage and flash at 1 second interval When the battery voltage is as low as the shutdown voltage the unit automaticall...

Страница 12: ...When the Stim Unit is in the pause state press the Electrical Stimulation Intensity Adjustment Button to activate the electrical stimulation intensity press the up button to increase the electrical st...

Страница 13: ...stool relax the affected leg and bend naturally Place the Stim Unit to correct position under the knee 4 1 3 3 Power on and operate Turn on Press and hold the switch mode composite button for 2 secon...

Страница 14: ...it receives a reply from the Stim Unit and pauses the walk Continue The App sends the gait mode start at which time the current intensity given in the parameter setting is the current intensity before...

Страница 15: ...lation stop the electrical stimulation by the release and continue to hold the electrical stimulation for continuous output The parameters that can be adjusted are the electrical stimulation intensity...

Страница 16: ...pp sends an evaluation mode to start the command to the Stim Unit receives the Stim Unit reply and enters the evaluation work interface Stop After long press and stop for 2s send the evaluation mode s...

Страница 17: ...in Care Before and after using the device please examine the skin it is normal that skin slightly turns red which shows faster blood flow Allergy Prevention Advice Do not position on the skin with mak...

Страница 18: ...ng to use the device Remember to regularly ventilate the skin covered by the Stim Unit Q4 The unit display battery icon is turned off after 1 second interval flashing A This indicates that the Stim Un...

Страница 19: ...ed 8 5 Product configuration Stim Unit 1pc Power Adapter 1set User manual 1pc 9 Product classification a Classified by type of electric shock internal power supply b The application part is classified...

Страница 20: ...epaired by unauthorized repair store 3 When asking for warranty service please take with the warranty card It is charged according to the stipulation of the repair service of the warranty Please conta...

Страница 21: ...trasse 80 D 20537 Hamburg Germany US Agent Yang LingYun 605 Sherwood Forest Place Cary North Carolina Cary NC 27519 Phone 919 4751044 Ext Email Usagency 126 Com Date 2018 12 17 Doc No XFT 2001EA E Ver...

Отзывы: