background image

 

Wireless  LED  HeadLight  System  (L2SWE)                                                                                                  www.xenosys.co.kr 

 

1. The battery pack for L2SWE is Lithium-Polymer rechargeable one. 

2.  When  transporting  only  battery,  make  sure  to  keep  it  in  soft  case.  Do  not  make  any 

conductor to connect the battery charging point. 

3.  The  battery  operating  time  may  shorten  using  in  the  cold  place  (below  (-10℃)  even  fully 

charged  one.  In  that  case,  warm  up  the  battery  within  the  pocket  before  using  L2SW.  Do 

not heat up the battery with microwave or any heater. 

4. Use only the charger which is provided by Xenosys. 

5. Charger battery in the range of 0℃  ~ 40℃. It may take longer to charger the battery out of 

this temperature range. 

6.  Provided  battery  can  be  charged  and  discharged  about  300  times  at  room  temperature. 

Lithium-ion battery is expendables, thus if the operating time is shortened after full charge, 

we recommend exchanging new battery.   

7.  Be  careful  on  benzene,  thinner,  alcohol  (volatile  &  flammable  fluid)  and  water.  If  those 

chemicals penetrate the battery, it may cause the trouble in circuit and any accident. 

8.  If  there  is  any  swollen  battery,  spill,  emit,  or  fire,  stop  using  immediately  and  contact  the 

distributor (dealer) or the manufacturer. 

9. Never use any damaged battery and charger. 

10.  In  case  of  the  first  installation  or  unused  long  time,  then  charge  the  battery  fully  before 

using. 

 

  Warning for Charging 

1.  Do  not  connect  any  metallic  objects  between  the  cradle  connector  and  battery.  This  may 

cause the defective on the charger and charger cradle.   

2. Use only the charger which is provided by Xenosys on the phone jack. This may cause the 

defective on the charger and charger cradle. 

  Warning for Disposal 

1.

 

In  case  of  careless  disposal,  it  may  pollute  the  environment,  so  make  sure  to  separate 

collection. 

Содержание L2SWE

Страница 1: ...s LED HeadLight System User Manual Optical Device Medical Device System Engineering XENOSYS CO LTD 2 Venture ro 100 Beon gil Yeonsu gu Incheon 22013 Korea TEL 82 32 875 9811 2 FAX 82 32 875 9813 www x...

Страница 2: ...e of Contents INTRODUCTION 3 SAFETY PRECAUTIONS 4 FEATURES 7 COMPONENTS 8 FUNCTIONS 9 OPERATING INSTRUCTION 12 BASIC TROUBLESHOOTING 16 SPECIFICATIONS 17 ELECTROMAGNETIC COMPATIBILITY 22 TRANSPORTATIO...

Страница 3: ...usting the illumination area adjustment knob This product is the medical device and KFDA approval registration is described like below table Registration No 1909 Certificate No Gyeongin Jesin 07 86 Pr...

Страница 4: ...o avoid the humidity and water 4 Do not use in the presence of a flammable anesthetic mixture with air oxygen or nitrous oxide Warning for Power Unit 1 Do not use any undesignated DC power supply exce...

Страница 5: ...se discoloration 7 Avoid the reach of children or pets 8 Maintain the following conditions for storing the product Environment for Storage Transportation Ambient Temperature 10 40 14 104 Relative Humi...

Страница 6: ...s thus if the operating time is shortened after full charge we recommend exchanging new battery 7 Be careful on benzene thinner alcohol volatile flammable fluid and water If those chemicals penetrate...

Страница 7: ...nput 100 240 V 50 60 Hz 0 3 A Output 5 V 2 2A 2 Charger Input DC5V 2 2A Output DC4 2V 1 0A 3 Battery Pack BTPWL DC3 7V 620mAh Lithium Polymer 4 Battery Pack BTPWS DC3 7V 380mAh Lithium Polymer 3 Prote...

Страница 8: ...sys co kr 8 COMPONENTS 1 Basic Components Main Body LED Control Unit LCUWE Battery Pack BTPWL 2 pcs BTPWS 2 pcs Charger LCHW Medical DC Power Supply FW8000MUSB05 USB Power Cable USBC KOIKO Mini Driver...

Страница 9: ...Refer to page 10 2 2 Charger LCHW Refer to page 11 1 Medical DC Power Supply FW8000MUSB 05 Refer to page 11 1 USB Power Cable USBC 1 KOIKO Mini Driver DRV18 1 Universal Clip 1 FUNCTIONS 1 Main Body L...

Страница 10: ...ersal clip or Xenosys TTL frame or loupe fix How to install Insert or temporarily assemble the M2 flat head bolt into the fix screw hole at the center of the fixing bracket shown in the left figure se...

Страница 11: ...Connector Jack Connector jack between DC5V power supply and the plug Charging Terminal The charging will start when the battery has been put on the charger and connect DC5V power supply connected to...

Страница 12: ...charge them fully before the first time usage Connect power supply charger to supply the power If the battery is fully charged it will show green light on charger LED Assemble the control unit fully...

Страница 13: ...water or electrolyte flow into the main body 6 In case of malfunctioning stop using immediately and contact the dealer 7 Do not disassemble repair or modify the equipment by unauthorized personnel Ca...

Страница 14: ...just the intensity change switch for desired brightness Below table will show you the operating time brightness lx WD400mm according to each step LED Lens Module LLMW Step Brightness lx WD400mm BTPWS...

Страница 15: ...ame by using universal clip E How to disassemble the battery pack 3 Sit on the battery pack to fit pogo pin of main body to the battery output terminal of battery pack or disassemble the battery pack...

Страница 16: ...urn off Sit on battery to the charger to check remaining value Re charge the battery The light intensity is not enough Check the intensity is properly adjusted Adjust the intensity step Discharging ti...

Страница 17: ...L2SWE www xenosys co kr 17 SPECIFICATIONS 1 Size and Weight A Main Body LED Control Unit LCUWE Size Refer to below photo Weight 19 g B Battery Pack BTPWS BTPWL Size Refer to below photo Weight BTPWS 1...

Страница 18: ...www xenosys co kr 18 C Charger LCHW Size Refer to below photo Weight 24 g D Medical DC Power Supply FW8000MUSB 05 Size 31 L x41 W x77 H mm Weight 81 g The purpose of green LED is to indicate the powe...

Страница 19: ...Wireless LED HeadLight System L2SWE www xenosys co kr 19 E USB Power Cable USBC Size 2m Weight 44 g...

Страница 20: ...minutes 3rd step mode Maximum Intensity Operating Time BTPWL Approx 8 hours 30 minutes 1st step mode Minimum Intensity Approx 3 hours 50 minutes 2nd step mode Approx 2 hours 30 minutes 3rd step mode...

Страница 21: ...D HeadLight System L2SWE www xenosys co kr 21 C Package External Figures Weight External Figures W228 x H138 x D80 mm Total Weight 550 g 3 Labels A Outer Carton Box B LED Control Unit C Battery Pack D...

Страница 22: ...vice is powered on you can find out if it generates electromagnetic interruption or not Recommend to choose one and more ways to solve the trouble in the following ways 1 Relocate the equipment receiv...

Страница 23: ...e as follows 2 Do not pull out twist nor bend excessively the power cable 3 Wind up the power cord loosely and store it by using the rubber bandage or cable tie 4 Do not expose the main body to excess...

Страница 24: ...untries to the extent that there is no special warranty rule 2 All services warranty should be provided by Xenosys Co Ltd or authorized personnel from Xenosys Co Ltd 3 Warranty becomes effective from...

Страница 25: ...ufacturer or distributor Then the manufacturer or distributor will report details with RMA No from your own generated and related reference which will make easier to track 11 There are no user service...

Страница 26: ...ys co kr 26 Revision List Version No Release Date Revision Version 1 0 January 29 2021 First version XENOSYS CO LTD 2 Venture ro 100 Beon gil Yeonsu gu Incheon 22013 Korea TEL 032 875 9811 2 FAX 032 8...

Отзывы: