INITIAL OPERATION
4.2
WATER QUALITy
• Do not use unfilterd and unsterilized main water!
Main water must be filtered with sterile filter with pore
size ≤ 0.22 μm and subsequently sterilized to avoid a
germ borne contamination.
• For filling the system, approx. 1 liter of sterile water is
necessary.
4.3
CONNECTINg THE DELTASTREAM HEATER COOLER
CAUTION
Risk due to water coming into
contact with electricity.
Do not connect the unit to the mains
until it has been filled with water.
CAUTION
Risk due to water coming into con-
tact with electricity.
Water can conduct electricity.
If the water overflows when the unit is
being filled, it must be dried thoroughly
and be completely dry before it is con-
nected to the mains and switched on.
When filling the unit for the first time,
please follow the instructions on disinfect-
ing (see “Maybe some of the listed prod-
ucts are not available in your country. Take
care about the availability!” on page 27).
The system is a closed circuit and the
water line should not be disconnected
in the operating room. Ensure that the
connection of the hose to the oxygenator
has already taken place before the system
is placed in the operating room.
4.3.1
FILLINg THE SySTEM WITH WATER
• Unsrew the cap on the water filler (2) without con-
taminating the interior of the unit.
Wear gloves when doing so.
• Be careful not to lose the seal on the cap.
• Add sterile water, for example by placing a funnel
in the water tank.
• We recommend using a sieve to prevent foreign
material entering the water circuit.
• Watch the water level gage (5) during filling. The
water level should be just under the MAX mark
after filling.
• Screw the cap on the water filler tightly until it is
sealed.
Figure 1:
5
2
Filling the system with water