33
NMS 450X
m
en
u
en
ter
NMS 450X
NMS 450X
m
en
u
en
ter
NMS 450X
NMS 450X
m
en
u
en
ter
NMS 450X
NMS 450X
m
en
u
en
ter
NMS 450X
5.1.3) Neuromuscular Transmission Monitoring Mode
(NMT Mode) (NMS450X only)
•
Insert the NMBA Cable.
The following screen should appear on the display.
•
Ensure that the device is in ‘TOF’ mode.
•
Short-circuit the red and black electrode connectors.
•
Use the adjusting wheel and increase the current to 80mA.
•
Press the ‘play/pause’ button while shaking the accelerometer.
The NMS450 should respond as follows:
•
The LED should flash GREEN in accordance with the four stimulations.
•
Each stimulation should be accompanied by an audible ‘beep’.
•
In ‘Diagnostic window’ four bars of different heights should indicate that the
accelerometer detected movement.
•
Monitor the actual current delivered to ensure that the warning sign does not appear.
•
Separate the red and black electrode connectors to cause an open circuit between
them.
•
Press the ‘play/pause’ button.
The following screen should appear on the display.
•
No audible feedback should be heard.
•
If the STIMPOD malfunctions in any one of these performance tests, it should
be checked by the relevant technical department in accordance with the test
instructions in the Technical Service Manual.
•
Equipment may only be repaired by the manufacturer or by an organisation expressly
authorised by the manufacturer.
•
Equipment does not require regular calibration.
Open Circuit
Detected
Open Circuit
Detected