3.10) Tetanus (TET)
Defaults:
50Hz (adjustable to 100Hz)
The Tetanus stimulation comprises series of 200 microseconds square wave pulses
repeated at a 50Hz or 100Hz repetition rate or frequency.
Selecting TET Mode:
•
Insert the NMBA cable with the red and black electrode clips as well as the tri-axial
accelerometer.
•
Press the ‘Mode’ button until ‘TET’ is shown on the display.
Stimulate/Stop:
Start stimulation by holding the ‘Play/Pause’ button down. Stop stimulation by releasing
the ‘Play/Pause’ button.
3.9) Twitch (TWI)
Defaults:
Repeat at 2Hz
Adjustable:1Hz, 2Hz and 5Hz
The Twitch stimulation comprises a 200 microseconds square wave pulse. If the ‘Play/
Pause’ button is pressed the twitch will repeat at the selected frequency.
Selecting TWI Mode:
•
Insert the NMBA cable with the red and black electrode clips as well as the tri-axial
accelerometer.
•
Press the ‘Mode’ button until ‘TWI’ is shown on the display.
Stimulate/Stop:
Start stimulation by pressing the ‘Play/Pause’ button. Stop stimulation by pressing the
‘Play/Pause’ button again.
Frequency adjustment:
Press the ‘Hz’ button to toggle through the frequencies.