Resetting the Power Supply
M370078-01
3-39
3
Resetting the Power Supply
The reset is used to clear the parameters to the factory default values.
Soft Reset
The soft reset is used to set the parameters (see Table 3-9) to the default
values, but it does not reset the calibration constants
To perform a soft reset:
1. Turn the power supply to Off then On.
When the unit is powering on,
8888
8888
is displayed on the output
voltage and current displays.
2. Press the key combination of both OUTPUT ENABLE Main button
and OUTPUT ENABLE Aux button and hold continuously for 3
seconds in order to execute.
If the flash fail alarm triggers, it is also possible to reset the power
supply.
3. Once a reset has been triggered, the output voltage display shows
P5U
CLr
for 1 second.
The model number will be displayed for 1 second and then return to
normal mode.
After executing a reset, the power supply’s settings are restored to the
default settings but retain the calibration data. Table 3-9 lists the default
settings.
Table 3-9
Power Supply Default Settings
Parameter
Setting
Address
1
Data rate (Kbps)
9.6
Communication mode
RS-232
Vout setting
0 V
Iout setting
0 A
Output
Off
AUX Out
Off
Trigger
Off
Ext. On/Off Polarity
Negative
Auto start mode
Off
AUX Auto start mode
Off
OVP
Maximum
Содержание Watt Series Programmable DC Power Supply XG 850
Страница 2: ......
Страница 3: ...XG 850 Watt Series Programmable DC Power Supply Operating Manual...
Страница 6: ......
Страница 18: ...xvi...
Страница 30: ...1 8...
Страница 52: ...Local Operation 3 8 M370078 01 Figure 3 3 Front Panel Menu System...
Страница 82: ...Local Operation 3 38 M370078 01 The SCPI command s for these instructions are SYSTem channel s REMote STATe...
Страница 86: ...Local Operation 3 42 M370078 01 Figure 3 5 Split Supply Operation V V Common Power Supply Power Supply Load...
Страница 126: ...Analog Programming APG and Isolated Analog Programming ISOL 4 34 M370078 01...
Страница 198: ...5 72...
Страница 262: ...A 36...
Страница 284: ...IX 4...
Страница 285: ......