CALIBRATION

M9B-HFS INTERFACE

4-4

Release 1.0 (97/10)

4.3.3 

Current Readback Calibration Procedure

1.  Connect the current shunt to the power supply as shown in Section  4.3.1 Current Calibration Setup.

Connecting a DVM is optional.

2.  Activate calibration mode by sending software command CMODE ON to the power supply.

3.  Send command IRLO; VRLO to the power supply. Wait for the supply to settle. Measure and record the

voltage output shown on the external meter. Send IRLO again.

4.  Send command IRHI; VRHI to the supply. Wait for the supply to settle. Measure and record the output

voltage as shown on the external meter. Send IRHI again.

5.  Calculate IRLO and IRHI from the voltages taken from the external meter and the shunt resistance.

I=V/R.

6.  Send the command IRDAT <ilo>,<ihi> where <ilo> and <ihi> are the current values obtained by sending

the commands IRLO and IRHI to the supply. When the power supply is calibrated, the low to high
current readback calibration values are stored as constants. Refer to the current readback accuracy
specification in Appendix A.

7.  Use commands ISET and IOUT? to confirm that the calibration was successful and that linearity is

observed. Refer to the voltage readback accuracy specification in Appendix A.

8.  Turn off calibration mode by sending the command CMODE

 OFF to the power supply.

4.4  Over Voltage Protection (OVP) Calibration

We recommend that you perform OVP calibration every six months. Connecting a digital voltmeter as in Section
4.2.1 Voltage Calibration Setup is optional.

1.  Disconnect any load from the supply.

2.  Ensure that jumper J66 on the interface PCB is installed to enable remote OVP calibration. (Jumper J66 is

installed at the factory).  See 

Figure  2.7-1 for the location of jumper J66.

3.  Activate calibration mode by sending command CMODE ON to the supply.

4.  Send the command OVCAL to the supply. The ADDR LED will light during OVP calibration. Calibration

is complete when the ADDR LED turns off. This may take a few minutes.

5.  Use the OVSET, OVSET?, and VSET commands to trip the OVP level, confirming that the calibration

was successful. When you trip the OVP level, the red OVP LED will light and the voltage will drop to
zero. Send the command RST to clear the OVP condition. Refer to the OVP program accuracy
specification in Appendix A.

6.  Turn off calibration mode by sending the command CMODE

 OFF to the supply.

Содержание Internal M9B-HFS Series

Страница 1: ...TM 9BHF XFXN Technical Manual for Internal M9B HFS Serial Interface for XFR Series Power Supply...

Страница 2: ...contained in this document is confidential and is the exclusive property of Xantrex Technology Inc It may not be disclosed to any person without the express written consent of Xantrex Technology Inc P...

Страница 3: ...Details the inspection of the interface card and basic setup Configuration procedures Section 3 M9B HFS Operation Lists the complete command set status registers and error codes Section 4 Calibration...

Страница 4: ...MANUAL CHANGES M9B HFS INTERFACE ii Release 1 0 97 10 MANUAL CHANGES There are no changes at this time...

Страница 5: ...Fiber Optic Line 2 5 2 4 1 Fiber Optic Cable 2 5 2 4 2 Fiber Optic Serial Interface 2 5 2 5 Power ON Conditions 2 6 2 5 1 Default Conditions 2 6 2 5 2 Remote Enable REN 2 6 2 5 3 Local Lockout LLO 2...

Страница 6: ...3 4 3 3 Current Readback Calibration Procedure 4 4 4 4 Over Voltage Protection OVP Calibration 4 4 5 MAINTENANCE 5 1 Introduction 5 1 5 2 Troubleshooting 5 1 5 2 1 Diagnostic LEDs 5 1 5 3 Replaceable...

Страница 7: ...ters 3 13 Table 3 5 1 Error Codes 3 14 Table 5 3 1 M9B HFS Interface PCB Replaceable Parts 5 2 Table 5 3 2 Fiber Optic Connection Parts 5 4 Table A 1 Electrical Specifications for the 1200 Watt Suppli...

Страница 8: ......

Страница 9: ...mpact of noise on communications to the power supply and improves signal quality The M9B HFS features an extensive command set and several user programmable features In addition to fiber optic cable a...

Страница 10: ...D status signals error address remote local operation and over voltage protection Foldback in CV or CC mode with reset Software calibration Software communications flow control Asynchronous full duple...

Страница 11: ...wer supply On first receiving your unit perform a quick physical check 1 Ensure each package contains a power supply with its M9B HFS Interface board installed and manuals for the power supply and the...

Страница 12: ...ep Procedure References 1 Set the back panel S1 Communication Switch for operation See Section 2 3 S1 Switch Communication Settings for information on setting the Baud Rate and Flow control for the in...

Страница 13: ...Panel J7 User Signal Connector Transmitter LED Fiber Optic Receiver Port Remote Programming LED REM Addressed LED Error LED Fiber Optic Transmitter Port Return to Local Button LOCAL Addressed LED ADR...

Страница 14: ...2 3 1 S1 Switch Settings for Baud Rate Selection BAUD Rate B3 B2 B1 9600 1 1 1 4800 0 1 1 2400 1 0 1 1200 0 0 1 600 1 1 0 300 0 1 0 150 1 0 0 75 0 0 0 2 3 2 Flow Control Flow control signals regulate...

Страница 15: ...r HFBR 1521 Fiber optic transmitter HFBR 2521 You need two fiber optic lines one for transmitting and one for receiving These cables connect to the interface at the Rx and Tx ports on the rear panel s...

Страница 16: ...see models IMAX supply s current limit OVP Trip Voltage Model VMAX 10 OVSET voltage limit 10 Delay 0 5 seconds DLY 0 5S Foldback Protection OFF FOLD OFF Output ON OUT ON Hold OFF HOLD OFF Unmask NONE...

Страница 17: ...return the power supply to local operation by using the REN software command or the front panel LOCAL button control of the OVP trip level switches to the front panel OVP potentiometer The default OV...

Страница 18: ...tion closed default See Section 2 7 3 open See Section 2 7 3 Note All other jumpers are not user selectable LED INDICATORS CR97 Red Diagnostic LED Bus error or soft restart on Slave circuitry CR163 Re...

Страница 19: ...you never plan to control the power supply from the front panel Otherwise the software commands and the front panel return to local LOCAL button will offer you greater flexibility in controlling your...

Страница 20: ...o locate the connector refer to the M9B HFS interface rear panel drawing in Figure 2 2 2 Pin 1 is located in the top right position of connector J7 when you face the rear panel See also Section 3 2 Co...

Страница 21: ...text full capitals 3 1 2 Command Format and Parameters The device dependent language for the M9B HFS Interface consists of commands and parameters A command is a one word code which either gives inst...

Страница 22: ...and UNMASK commands in the command reference for use of the ALL and NONE parameters See registers in Section 3 4 state The state of a binary condition 1 ON 0 OFF Floating Point Number float Variables...

Страница 23: ...detailed information about each command and its use Table 3 2 1 Programming Commands Command Description AUXA Selects the state of the AUXA output signal on the J7 7 connector AUXB Selects the state...

Страница 24: ...s present output voltage setting Table 3 2 3 Calibration Commands Command Description CMODE Places the supply into calibration mode IDATA Calculates the slope and intercept for current programming IH...

Страница 25: ...operating conditions as listed in the status register AUXA state Parameters 1 ON 0 OFF Controls the AUXA output signal level at rear panel connector J7 7 Active low Initial value AUXA 0 AUXA Asks for...

Страница 26: ...he accumulated status and fault registers If the ERR bit has been masked using the MASK command then the ERR bit in the registers will not be set Once an error is detected the remaining portion of the...

Страница 27: ...is enabled so that all voltage and current settings which would normally be implemented by the supply are held until a TRG trigger command is received Initial value HOLD 0 OFF HOLD Asks for the prese...

Страница 28: ...ameters Ilo Ihi Calculates and records the slope and offset for readback voltage using IRLO and IRHI data Set CMODE ON before using this command See also the calibration procedures in Section 4 IRHI T...

Страница 29: ...1 is received OUT OFF or OUT 0 also asserts the isolation signal on the rear panel J7 connector line 3 You can use the J7 3 user signal to trip external relays to isolate the power supply from the lo...

Страница 30: ...the total bit weights for the operating conditions as listed in the status register TRG Implements programmed voltage and current settings which had been in hold mode The supply operates with previou...

Страница 31: ...ommand See also the calibration procedures in Section 4 VMAX volt Sets an upper soft limit on the supply s programmed output voltage If the soft limit is exceeded or if the soft limit value is lower t...

Страница 32: ...n 4 VSET volt Parameter voltage Sets the power supply s output voltage in volts default or in millivolts This programmed voltage is the actual output in CV constant voltage mode or the voltage limit i...

Страница 33: ...ond to each register condition You can select conditions you want to monitor in the fault and status register by using the MASK and UNMASK commands Table 3 4 1 Accumulated Status Status and Fault Regi...

Страница 34: ...y placed word number separator or terminator was received Example OFF VOUT 6V MASK ERR 5 Number Out of Range The numerical value for the command was outside of the allowed range 6 Attempt to Exceed So...

Страница 35: ...y s main assembly You may need to recalibrate the interface whenever you install a new interface board or if the unit falls out of specification due to component aging drifts You can calibrate the M9B...

Страница 36: ...g the command CMODE OFF to the supply 4 2 3 Voltage Readback Calibration Procedure 1 Set up for calibration as in Section 4 2 1 Voltage Calibration Setup 2 Send the software command CMODE ON to activa...

Страница 37: ...ON to the power supply 3 Send commands ILO VLO to the power supply Measure and record the output voltage shown on the external voltmeter 4 Send commands IHI VHI to the supply then measure and record t...

Страница 38: ...IOUT to confirm that the calibration was successful and that linearity is observed Refer to the voltage readback accuracy specification in Appendix A 8 Turn off calibration mode by sending the comman...

Страница 39: ...cuitry caused a transparent restart An internal bus error will also attempt a transparent restart A lighted red COP LED at CR97 indicates that noise in the slave processor circuitry caused a transpare...

Страница 40: ...CB 103F 16 C61 47nF 50V X7R 10 5 0mm Ceramic Radial Capacitor CB 473F 06 C182 188 189 3 3nF 1KV Z5U 80 to 20 6 5mm Cer Rad Capacitor CC 332G 67 CR9 24 36 45 47 56 60 90 90 111 147 1N4148 UR D035 75V 3...

Страница 41: ...135 194 202 8 pin Machined Socket 0 3 MC M308 IC U81 16 bit Resolution DAC UD 1600 JP For U81 24 pin Machined Socket 0 6 MC M624 IC U94 171 Quad Pos NAND Gates with Schmitt Trig UH C132 U95 96 116 20...

Страница 42: ...onnect fiber optic line to it using a fiber to RS232 serial data converter Table 5 3 2 shows the part number and supplier of a fiber optic to RS232 serial data converter Standard fiber optic cable ass...

Страница 43: ...V 1 0mA 1 3mV 0 7mA Readback Accuracy1 Voltage Current 30mV 0 12 500mA 0 1 75mV 0 12 250mA 0 1 75mV 0 2 165mA 0 15 75mV 0 3 140mA 0 15 150mV 0 25 120mA 0 1 Table A 2 Electrical Specifications for XFR...

Страница 44: ...1 3mV 1 6mA Readback Accuracy1 Voltage Current 10mV 0 12 300mA 0 15 75mV 0 12 165mA 0 15 75mV 0 2 120mA 0 15 75mV 0 3 110mA 0 15 150mV 0 3 80mA 0 1 Table A 4 Electrical Specifications for 2800 Watt Su...

Страница 45: ...M9B HFS INTERFACE APPENDIX B Release 1 0 97 10 B 1 B ASSEMBLY SCHEMATIC M9B HFS OPTION BOARD XS 9BHF...

Страница 46: ......

Страница 47: ......

Страница 48: ...Xantrex Technology Inc 8999 Nelson Way Burnaby B C Canada V5A 4B5 Tel 604 422 8595 Fax 604 420 1591 Toll Free North America 1 800 667 8422 e mail info xantrex com http www xantrex com...

Отзывы: