S P 6 0 S E R I E S
S P H E R E S P E C T R O P H O T O M E T E R
9
Analyze Mode
The Analyze mode allows you to take measurements and
display the resulting color data (actual and difference) without
storing the data in the instrument. Measured data can be
transmitted from the RS-232 port.
Compare Mode
The Compare mode is a quick method for comparing
measurements without storing the data. After entering the
mode, the first measurement is set as the standard and each
measurement thereafter is compared to it. The standard can be
changed by pressing the Enter key
#
and remeasuring.
Strength Mode
The Strength mode is used when a colorant is added to a
sample to match a standard. The strength method and predicted
mode used are selected in the instrument configuration.
Opacity Mode
The Opacity mode is used for making multiple measurement
calculations to determine contrast ratio or percent opacity. Each
measurement requires three readings (over white, over black,
and a measurement of the white backing). The final data can be
stored and displayed either as over white, over black, or color at
full hide (equivalent color @ 100% opacity) as determined by
configuration settings.
•
Data Storage Information
– When in QA, Strength, or
Opacity mode, this area displays the project, standard, and
sample information associated with stored data. Repeatedly
pressing the Enter key
#
when
Proj ##
or
Std ##
is
highlighted pages through the available projects or
standards stored in the instrument (or hold down the Enter
key
#
to access a specific number from the editor dialog).
Pressing the Enter key
#
when
Sample
is highlighted
activates the sample Database Tools menu. Pressing the
Enter key
#
when a project name or a standard name is
highlighted displays the setup information for the selected
item. When the instrument is in storage mode, repeatedly
Data Storage
Information
Color Data
Color Data
Parameters
P r o j 1 : C a r t o n s
S t d 1 : R e d S a m p l e
S a m p l e :
# 1 0 1 1 : 2 3
L*a*b*
D65/10
SPIN
....
.....
.
∆
L* +0.05
L* 88.25
∆
a* —0.03
a* —4.71
∆
b* —0.14
b*+36.64
∆
E* 0.16