RapidMatch
™
XI Spectrophotometer
11
a calibration must be performed. No measurements can be taken until the
calibration is completed.
NOTE: Make sure to use the calibration reference supplied with the
instrument for calibrating. Do not substitute this reference with a
reference from another instrument. The serial number on the
reference should match the reference serial number displayed on
the instrument screen during calibration.
•
The white tile in the calibration reference is dramatically affected by
smudge marks, dust, and finger prints.
Cleaning the Calibration Reference
The white tile and effect tile in the calibration reference should be
cleaned using a mild soap and warm water solution, thoroughly rinsed
with warm water, and wiped dry with a clean, lint-free cloth. You must let
the reference dry completely before taking a calibration measurement.
•
Do not move instrument while taking a calibration measurement. If
motion is detected, an error message will be displayed and calibration
aborted.
•
When in calibration mode, swipe the screen to the left for additional
information on positioning.
1.
Select the Calibration option from the Settings menu.
Note
: If
“Calibration required” message appears, advance to Step 2.
2.
Tap Calibrate to continue with the calibration.
3.
Lay the calibration reference flat. Position the instrument on the
calibration reference over the white tile (1) so that the three pressure
sensors around the measurement port align with the slots in the
calibration reference, and the white line aligns with the reference marks.
Содержание RapidMatch XI