S p e c t r a L i g h t Q C
26
5.
Select desired option or value for the user setting.
To select option or value:
•
Use the up (▲) and down (▼) arrow swit
ches to highlight the option or select a value.
•
Press the
OK
switch to save the selected option/value.
6.
When finished with the user settings, press the
EXIT
switch to exit the User Settings screen.
Display System Information (INFO)
The System Info screen is used to view the current system information (i.e., versions, calibration status, daisy
device serial number, etc.). A system log file can also be viewed from this screen.
1.
From the main menu of the SpectraLight QC, press and hold the
MORE
switch until Setup appears in the
display screen.
2.
Press the
INFO
switch to access the System Info screen.
3.
Use the up (
▲
) and down (
▼
)
arrow switches to view all system information.
4.
The system information is displayed. Press the
LOG
switch to view the last 1024 log file entries.
5.
When finished, press the
EXIT
switch to exit the System Info screen.
NOTE:
The TEST switch is used to run a lamp test on the system. The test results are stored in the unit and can
be output at a later time by the software application to generate a test report.