CALIBRATION MODE 

The calibration consists of a ceramic plaque for white measurements and a trap opening for black 
measurements. 
A calibration should be performed when prompted for by the instrument or when desired.  
Refer to Cleaning section in the Appendices for information on cleaning the optics area and 
references. 

NOTE: 

Make sure to use the calibration reference supplied with the instrument for calibrating. 

Do not substitute this reference with a reference from another instrument. The serial number 
on the reference should match the reference (plaque) serial number on the instrument.

 

Calibration Notes

 

 

Dirt or dust in the aperture area will cause an inaccurate calibration reading. Refer to the 
Appendices for optics cleaning procedure. 

 

The 

white ceramic plaque in the calibration reference is dramatically affected by 

smudge marks, dust, and finger prints

. Refer to Appendices for calibration reference 

cleaning procedures. 

 

The 

black trap should be cleaned periodically to remove any dust or 

contamination

. Refer to Appendices for black trap cleaning procedures. 

 

Do not move instrument while taking a calibration measurement

. If motion is 

detected, calibration will be aborted. 

 

Important:

 If the lamp signal in your instrument reaches a level that is 

50% below the factory level stored, the warning symbol shown at the right 
appears after a calibration is completed. This indicates that the illumination 
level is getting low and you should have your instrument serviced soon. 

 

White and Black Calibration

 

Procedure 

1.

 

From the Main screen, use the 

Navigation 

control to move the highlight square to the 

Calibration icon. 

 

2.

 

Tap the 

Select

 control to access the calibration screen. 

Calibration Required appears if calibration is needed. If calibration is not currently needed, the 
time remaining before the next calibration along with the white reference (plaque) serial 
number appears. To exit calibration mode without calibrating, select the Exit icon (

x

) in the 

Progress column. 

Содержание Ci61

Страница 1: ...Ci61 Spectrophotometer User Guide...

Страница 2: ......

Страница 3: ...and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a...

Страница 4: ...ications or failure to follow the maintenance procedures in X Rite s accompanying documentation or published specifications iii repair or service by anyone other than X Rite or its authorized represen...

Страница 5: ...AN AN OFFICER OF X RITE IS AUTHORIZED TO MAKE ANY WARRANTY IN ADDITION TO THE FOREGOING IN NO EVENT WILL X RITE BE LIABLE FOR ANY OF BUYER S MANUFACTURING COSTS OVERHEAD LOST PROFITS GOODWILL OTHER EX...

Страница 6: ...de 13 Configuration Mode 13 Display Screen Layout 13 Header Bar 13 Main Column 14 View Column 14 Content Area 14 Progress Column 14 Configuration Mode 15 Entering Configuration Mode 15 Measurement Tri...

Страница 7: ...5 Appendices 28 Service Information 28 Cleaning the Instrument 29 General Cleaning 29 Cleaning the Optics 29 Cleaning the Calibration Reference 30 Replacing the Battery Pack 32 Troubleshooting 33 Scre...

Страница 8: ...control for easy screen selections measure button and power on off Packaging Your instrument packaging should contain all the items listed below If any of these items are missing or damaged contact X...

Страница 9: ...nd around your wrist The strap should not be used to carry the instrument 1 Feed the small looped end of the strap around the post at the back of the instrument 2 Insert the wrist strap end through th...

Страница 10: ...agnostics test and displays a splash screen before the main menu screen appears Splash Screen Main Menu Locking the Navigation Control The power button is also used to lock the navigation control This...

Страница 11: ...ours for full charge A charged battery pack may eventually lose partial charge if not used for an extended amount of time You should charge the battery from time to time and store in a cool environmen...

Страница 12: ...detachable line cord in the AC adapter and plug the line cord into the wall receptacle Connecting the USB Cable IMPORTANT You must install the software before connecting the instrument to your compute...

Страница 13: ...ning the instrument off and then back on will reset their operation Measure Button The Measure button is located at the front of the instrument The button can be configured to initiate a measurement o...

Страница 14: ...me height as the item for the rest of the instrument s shoe to sit on Procedure 1 Clear the sample surface of any dirt dust or moisture 2 Position the target window over the sample to measure If possi...

Страница 15: ...alibration Mode The calibration mode is used to perform white reference and black trap calibrations Refer to the Calibration Mode section for information Configuration Mode The configuration mode is u...

Страница 16: ...t condition of the battery pack Indicates the battery pack is fully charged Indicates the battery pack has a sufficient charge for a substantial number of measurements Indicates the battery pack is lo...

Страница 17: ...navigation buttons to move the highlight focus to the desired configuration icon in the Options column NOTE The arrow icon or at the end of the Options column indicates that additional options are av...

Страница 18: ...gger a measurement Pressure Shoe closing is required to take a measurement Button Measure button is required to take a measurement Pressure and Button Both the read switch and measure button are requi...

Страница 19: ...rameter and then tap the Select button 4 Repeat steps 1 through 3 until the date and time are set 5 Tap the Right navigation button and highlight the checkmark icon in the Progress Column 6 Tap the Se...

Страница 20: ...Refer to Appendices for calibration reference cleaning procedures The black trap should be cleaned periodically to remove any dust or contamination Refer to Appendices for black trap cleaning procedur...

Страница 21: ...erence 5 Position the instrument target window over the white ceramic plaque 6 Press the instrument firmly to the shoe and hold steady until the display indicates that the reading is complete Release...

Страница 22: ...control to save the calibration and exit to the main menu 11 Store the calibration reference in a dry dust free area away from direct exposure to light NOTE If an error message appears during or after...

Страница 23: ...t timestamp if desired Refer to Renaming the Sample Name later in this section Entering Measure Mode 1 From the Main screen use the Navigation control to move the highlight square to the Measure mode...

Страница 24: ...or Down navigation buttons to move the highlight focus to the Project icon 2 Tap the Select or Right button to access the project selection area 3 Use the Up or Down navigation buttons to move the hi...

Страница 25: ...cter space and tap the Select button 9 Use the Up or Down navigation buttons as previously explained to set the character and tap the Select button 10 Continue until all characters for the name are se...

Страница 26: ...s Column 2 Tap the Select button and then tap the Left navigation button to move the highlight to the select option screen 3 Use the Up or Down navigation buttons to choose the Lock Project option Tap...

Страница 27: ...elect button 9 Use the Up or Down navigation buttons as previously explained to set the character and tap the Select button 10 Continue until all characters for the name are set 11 Tap the Right navig...

Страница 28: ...ugh the alphanumeric list for the first character of the name NOTE You can also use the navigation control to rapidly scroll through a list by touching the control between the arrows and circle left o...

Страница 29: ...g should be used when measuring non uniform samples textured materials and any dark or high chroma colors The following example has an average setting of 2 To measure using averaging 1 Select the proj...

Страница 30: ...the world You can contact us using one of the following methods To identify the X Rite service center nearest you please visit our web site at www xrite com For online help visit our web site www xrit...

Страница 31: ...water or mild cleaner NOTE DO NOT use any solvents to the clean the instrument this will cause damage to the cover Cleaning the Optics The optics should be cleaned once a week in a normal environment...

Страница 32: ...ve the ceramic plaque from the base 3 Clean the plaque using a mild soap and warm water solution Thoroughly rinse with warm water and wipe dry with a lint free cloth Do not use solvents or cleaners of...

Страница 33: ...your fingers and separating 2 Clean with clean dry air or wipe clean with a lint free cloth to remove any dust or contamination 3 After cleaning align the two arrows on the end before reassembling th...

Страница 34: ...s cover latch while lifting upwards on the cover to remove it from the instrument housing 3 Remove the old battery pack and dispose in a designated disposal location for recycling 4 Insert the new bat...

Страница 35: ...n procedure fails Calibration reference is dirty or damaged Clean the reference per procedure in Appendix or replace if damaged Instrument and software not communicating Interface cable not connected...

Страница 36: ...Max 0 50 E ab on any tile specular component included Short Term Repeatability 10 E ab on white ceramic standard deviation Lamp Life Approx 500 000 measurements Power Supply Removable Li ion battery p...

Страница 37: ......

Страница 38: ...adquarters X Rite Europe GmbH Althardstrasse 70 8105 Regensdorf Switzerland Phone 41 44 842 24 00 Fax 41 44 842 22 22 Asia Pacific Headquarters X Rite Asia Pacific Limited Suite 2801 28th Floor AXA To...

Отзывы: