Safety Points:
The Patient Weight Scale must be used in accordance with these instructions and in conjunction
with the instructions for the lifting device and the sling. Do not exceed the patient weight capacity
of the lifting device or sling.
The maximum lifting capacity (patient weight) of the Patient Weight Scale is 600 lb (272 kg).
Do not exceed the lifting capacity.
The Patient Weight Scale is intended to be operated entirely by a caregiver. The patient should
not perform any function relating to the control of the scale, sling or lift. Additional caregivers may
be required with certain patients or situations.
If the Patient Weight Scale becomes soiled or contaminated it can be cleaned following the
instructions found in
the “Maintenance” section of this guide.
Inspect the Patient Weight Scale carefully before each use. If structural fatigue, erratic readings
or other damage is visible or suspected discontinue use immediately.
Before attempting to weigh, the patient must be assessed by a qualified professional.
Always make sure the casters on the lift are unlocked during the weighing procedure.
Do not leave the patient unattended during the weighing procedure.
Operating the Patient Scale
1. Follow the instructions for operating the lift and sling.
2. Attach the empty sling to the sling bar securely.
3. Ensure the scale is clear of any obstructions, is hanging vertically and is free to pivot at both the
top and bottom connection points.
4. P
ress the “
Zero
” button to set the system to “
0.0
”. A “
0
” will pan across the screen for a couple
seconds.
Display will then read “
0.0
”. See “Fig. 1”
5. Press the
“
Lb/ Kg
” button to choose the desired mode. See “Fig. 1”
6. Remove the sling and following the sling instructions position the patient in the sling. Reattach
the sling to the sling bar ensuring the scale is hanging vertically and is free to pivot.
7. Carefully lift the patient and when the patient is fully supported by the lift and any motion has
stabilized, press the “
Operate
” button to display the patient weight. Do not touch the patient or
sling during this step.
Note:
Auto-Power-Down is set for 60 seconds to conserve the battery. To wake the system,
simply press the “
Operate
” button again.
8. When the weighing process is complete carefully lower the patient.
Fig. 1
Fig. 2
Indicates
negative weight
Patient weight
Weight mode
Lb/Kg, Operate, Zero
buttons