Logging start with clearing memory and logging mode selection
It is possible to clear memory only if logging is stopped. Modification of logging mode is
enabled only after memory is cleared. Clearing of memory is optional part of the above procedure
logging START. Notice: there is no way to "undelete" the cleared data!
If it is needed to clear the memory before logging start, first it is necessary
to select menu item logging START (symbol Strt) and confirm by pressing
ENTER key.
Notice: if low battery voltage is indicated, it is not enabled to start logging
nor to clear the memory
Now instrument offers possibility to clear the memory (ERAS). Default is
selection ERAS NO. If clearing of the memory is needed use the arrow keys
to select ERAS YES and confirm by pressing ENTER key..
Clearing procedure is in progress. It takes approximately 16 s and the
duration is indicated by displaying of already cleared memory in % (digits 0
to 100).
After memory is cleared it is possible to select one of three logging modes
(MODE) by using arrow keys (non cyclic logging is indicated by symbol
n.CYC, cyclic logging by symbol CYCL and manual logging by symbol
Man). If manual logging mode is selected, after confirmation by ENTER
key logging is active. In automatic logging mode (i.e. cyclic or non cyclic)
last applied storing interval is offered to confirm or modify. See above
procedure.
Logging stop from keyboard
Logging stop is enabled by confirmation of menu item STOP by pressing
ENTER key.
Reading STOP NO is displayed to enable logging stop. Use arrow keys to
modify to STOP YES. Logging is still in progress.
Logging is stopped after confirmation selection STOP YES by pressing
ENTER key. Instrument then returns to default mode of displaying actual
values.