Maintenance, cleaning, disposal
38
5.1.2 Changing the batteries
The batteries must be changed when the LoBat message
appears (approx. every 3 years). We recommend to have
the batteries exchanged by service personnel. If you want to
exchange the batteries by your own, proceed as follows:
Warning
When changing the batteries, do not disconnect the instru-
ment from the mains supply! The instrument requires a volt-
age supply to save the data. There is no risk to the user.
5 Stow the wires in the opening (3) of the measuring in-
strument again. Make sure that the wires are not
placed in front of the lamp and that they are not in the
way of the lamp module
6 Insert the new lamp module (lamp symbol must be
vertical) so that it locks into place and a click can be
heard
7 Connect the instrument to the mains supply again
and switch it on
8 Calibrate the instrument (see 4.3 C
ALIBRATION
). The
instrument is now ready for operation again.
1 Switch off the measuring instrument but do not dis-
connect it from the mains supply
2 Remove any cells from the instrument
3 Remove the lamp module and undo the plug-in con-
nection (see 5.1.1 C
HANGING
THE
LAMP
MODULE
)
4 Carefully turn the measuring instrument around
and
safely disconnect it
5 Remove the card with the short instructions
6 Undo the 5 screws on the underside of the instrument
using a Phillips screwdriver
Содержание Turb 550
Страница 1: ...ba41112e03 04 2003 Turb 550 Turb 550 IR ba41112e Operating Manual Turbidity Measuring Instrument...
Страница 8: ...Overview 8...
Страница 12: ...Safety 12...
Страница 16: ...Commissioning 16...
Страница 25: ...Operation 25 Note The flow through vessel can remain installed until the next calibration...
Страница 42: ...Maintenance cleaning disposal 42...
Страница 46: ...What to do if 46...
Страница 50: ...Accessories options 50...
Страница 54: ...Abbreviations and index 54...