Calibration
How to calibrate your measuring instrument:
1 Press the
key. The Ident display and the Cal dis-
play appear. The number 1000 appears in the lower
display line
2 Insert the 1000 NTU calibration standard in the cell
shaft
3 Align the cell and wait for a stable measured value
4 Press the
key.
The Store display flashes for approximately 3 sec-
onds. The number 1000 (NTU) appears in the upper
display line. The number 10.0 (NTU) appears in the
lower display line
5 Insert the 10.0 NTU calibration standard in the cell
shaft
6 Align the cell and wait for a stable measured value
Содержание Turb 550