Maintenance, cleaning, disposal

5.1

Maintenance

The TetraCon

®

700 SW conductivity measuring cell does not require 

maintenance.

5.2

Cleaning

Warning

Contact with the measuring solution or cleaning solution can 
endanger the user!Take protective measures suitable for the kind 
of measuring solution or cleaning solution (protective clothes, 
eye protectors etc.).

If the conductivity measuring cell is heavily contaminated, this can 
affect the measuring accuracy. Therefore, we recommend cleaning the 
conductivity measuring cell regularly according to visual checks. 
Especially before measuring low conductivity values, we recommend to 
clean the measuring cell thoroughly.

Cleaning

5.3

Disposal

We recommend to dispose of the conductivity measuring cell as 
electronic waste.

Contamination

 Cleaning agents

Reaction time at 
room temperature

Water-soluble 
substances

Tap water

Any

Slurry and loosely 
adhering dirt or 
biological films

Soft brush, warm 
tapwater with 
detergent

Any

Fats and oils

– Warm water and 

household 
detergent

– In the case of 

heavy 
contamination: 
Methylated spirits

– Any

– Maximum of 5 

minutes

Lime and hydroxide 
deposits

Acetic acid (10 %)

Any

Содержание TetraCon 700 SW

Страница 1: ...ba75538e01 04 2005 TetraCon 700 SW 19 Operating manual Conductivity measuring cell...

Страница 2: ...ces between this operating manual and your instrument Also we cannot guarantee that there are absolutely no errors in this manual Therefore we are sure you will understand that we cannot accept any le...

Страница 3: ...use 25 2 3 General safety instructions 26 3 Commissioning 27 3 1 Scope of delivery 27 3 2 Installation 27 4 Measuring Operation 29 4 1 Measuring 29 4 2 Application dependent settings 30 5 Maintenance...

Страница 4: ...TetraCon 700 SW 22 ba75538e01 04 2005...

Страница 5: ...ucture of the TetraCon 700 SW conductivity measuring cell Fig 1 1 Structure of the TetraCon 700 SW conductivity measuring cell 1 Shaft 2 Closing head with protective ring 3 Voltage electrodes 4 Temper...

Страница 6: ...from primary or secondary polarization effects which ensures a high degree of measuring accuracy The modern epoxy casting technique reduces the danger of breaking the conductivity measuring cell in ro...

Страница 7: ...e signal word e g Caution indicates the level of the danger Warning indicates instructions that must be followed precisely in order to prevent serious dangers to persons Caution indicates instructions...

Страница 8: ...conductivity measuring cell Protect the conductivity measuring cell particularly against frost or overheating Safe operation If safe operation is no longer possible the conductivity measuring cell mus...

Страница 9: ...ction to the measuring transmitter The TetraCon 700 SW connection cable has a cable plug to be connected to the Cond input socket of a measuring transmitter It is connected to the terminal strip of me...

Страница 10: ...Commissioning TetraCon 700 SW 28 ba75538e01 04 2005...

Страница 11: ...on If the gap is less than that the cell constant changes This leads to inexact measurement results Normally measuring transmitters have a correction function that can compensate for this influence Th...

Страница 12: ...due to the influence of the measuring environment e g of boundary fields In chapter 6 WHAT TO DO IF of this operating manual you will find the correct values to set for some products of the WTW acces...

Страница 13: ...measuring accuracy Therefore we recommend cleaning the conductivity measuring cell regularly according to visual checks Especially before measuring low conductivity values we recommend to clean the me...

Страница 14: ...y field not maintained The conductivity measuring cell must be surrounded by a gap of at least 5 cm at the base and sides when measuring at the electrodes Otherwise the cell constant will change see s...

Страница 15: ...lacement parts and accessories 33 ba75538e01 04 2005 7 Replacement parts and accessories Adapter Note Information on further accessories is given in the WTW catalog and in the Internet Model Order no...

Страница 16: ...Four electrodes measurement Temperature sensor Integrated NTC 30 30 k 25 C Dimensions Weight Material 59 5 160 37 5 223 40 approx 860 g without sensor connection cable Sensor head PVC epoxy filling ma...

Страница 17: ...delines and norms used Test certificates UL cUL Length 15 0 m special lengths on request Diameter 8 8 mm Smallest allowed bend radius Permanent bend 180 mm Short time bend 90 mm Connection Screw plug...

Страница 18: ...agram Pin assignment 7 pole screw plug Shield Thermistor NTC Thermistor NTC Current electrode I1 ring Voltage electrode U1 Voltage electrode U2 Current electrode I2 ring 6 1 2 3 5 7 4 NTC 30 k Measuri...

Страница 19: ...to article 3 3 of the directive 97 23 EC pressure equipment directive 8 4 Characteristics when delivered Temperature measurement Cell constant Measuring medium 0 C 50 C Storage transport 5 C 65 C stor...

Отзывы: