StirrOx G

Measuring / Operation

7

ba25310d07

08/2018

4

Measuring / Operation

4.1

Calibration

4.2

Measuring

Depth of

immersion

Observe the minimum and maximum depth of immersion of the sensor. The 
temperature sensor / auxiliary electrode must be completely submersed for 
measurement.

CAUTION

The sensor is designed for immersion measurements in BOD bottles.
Do not expose the sensor to overpressure

 

(see chapter 7 T

ECHNICAL

 

DATA

), e.g. due to sealed installation in a pressurized vessel. Danger of 

damage to the sensor.

Use of the
integrated

stirrer

The sensor is equipped with an integrated stirring system for the incident flow 
of the membrane.

Pressing the On/Off key switches the stirrer on. Pressing once more switches 
the stirrer off again.

Measurement in

sample bottles

The displacement volume of the StirrOx G (5,6 ml) is dimensioned so that the 
sample bottle is closed free of air bubbles after inserting the sealing plug. If 
you use bottles other than the Karlsruhe bottles available at WTW (see 
chapter 8 W

EAR

 

PARTS

 

AND

 

ACCESSORIES

), use WHEATON bottles or bottles 

with NS 19 standard ground joint only. The plug should have a displacement 
volume of 5.6 to 8.6 ml.

For calibration, please refer to the operating manual of the mea-
suring instrument.

Min. approx. 49 mm

Max. 83 mm

Depth of immersion

Содержание Stirrox G

Страница 1: ...OPERATING MANUAL ba25310e07 08 2018 StirrOx G DISSOLVED OXYGEN SENSOR WITH INTEGRATED STIRRER OPERATING MANUAL...

Страница 2: ...2 ba25310e07 08 2018 StirrOx G Copyright 2018 Xylem Analytics Germany GmbH Printed in Germany...

Страница 3: ...7 4 2 Measuring 7 4 3 Storage 8 5 Maintenance cleaning replacement 8 5 1 General maintenance instructions 8 5 2 Outside cleaning 9 5 3 Changing the electrolyte and membrane cap 9 5 4 Cleaning the elec...

Страница 4: ...Contents StirrOx G 4 ba25310d07 08 2018...

Страница 5: ...ottles 1 Shaft housing 2 Connection head 3 On Off key for stirrer 4 Cable screw joint 5 Stirring paddle 6 WP ST membrane cap filled with electrolyte solution 7 Gold working electrode cathode 8 Insulat...

Страница 6: ...to the instrument or the environment 3 Commissioning Scope of delivery D O sensor StirrOx G filled with electrolyte solution and ready to operate Calibration and storage vessel OxiCal ST ZBK ST acces...

Страница 7: ...sensor is equipped with an integrated stirring system for the incident flow of the membrane Pressing the On Off key switches the stirrer on Pressing once more switches the stirrer off again Measuremen...

Страница 8: ...ables you to place the calibration and storage vessel with the sensor in an upright tilt resistant position on a table 5 Maintenance cleaning replacement 5 1 General maintenance instructions For your...

Страница 9: ...thoroughly rinse with deionized water and recalibrate if necessary 5 3 Changing the electrolyte and membrane cap CAUTION Before starting to work with the sensor please note the GENERAL MAINTENANCE IN...

Страница 10: ...brane head and electrolyte solution see section 5 6 Rinse the sensor head with deionized water Carefully rub and dry the counter electrode with a lint free paper towel Immerse the sensor head includin...

Страница 11: ...crew on the membrane head fingertight using a paper towel Excess electrolyte solution is forced out of the ventilation area Check the filling Inspect the face surface No air bubbles may be present wit...

Страница 12: ...rew off the membrane head Caution Electrolyte solution For disposal of the membrane head and electrolyte solution see section 5 6 Rinse the sensor head with deionized water Using the rough side of the...

Страница 13: ...the membrane cap Additionally you can prevent air bubbles by throwing the first filling away and refilling the membrane cap Rinse the sensor head with electrolyte solution Hold the sensor inclined and...

Страница 14: ...on after 2 minutes 5 6 Disposal CAUTION The ELY G electrolyte solution irritates the eyes and skin When dealing with the ELY G electrolyte solution observe the following points During working activiti...

Страница 15: ...ated after changing the electrolyte and membrane cap Contaminated electrodes or sensor toxification Clean the electrodes see section 5 4 Leak display appears Membrane cap not screwed on tight enough H...

Страница 16: ...r PEEK Thermistor housing VA steel 1 4571 Seals FPM Viton Stirring paddle VA steel 1 4571 Connection cable Lengths 2 m Diameter 6 mm Smallest allowed bend radius Fixed installation 50 mm Flexible use...

Страница 17: ...mended storing method in the calibration and storage vessel OxiCal ST moist Storage temperature 0 50 C 32 122 F Characteristics when delivered Zero signal 0 1 of the saturation value Response time of...

Страница 18: ...assignment Ue V I0 mA I1 mA I2 mA 5 0 4 0 38 189 6 0 4 8 35 147 7 0 5 8 32 137 8 0 6 7 29 121 9 0 7 7 28 108 10 0 8 6 27 100 11 0 9 5 27 95 12 0 10 4 27 89 AE GE HE 7 5 3 1 8 6 4 2 NTC 1 30 kW NTC 2 3...

Страница 19: ...electrode RL G 205 204 Polishing film SF 300 203 680 Accessory kit comprising 3 exchange membrane caps WP ST Electrolyte solution ELY G Cleaning solution RL G Polishing strip SF 300 ZBK ST 202 710 St...

Страница 20: ...Wear parts and accessories StirrOx G 20 ba25310d07 08 2018...

Страница 21: ......

Страница 22: ...buildings factories and farms In more than 150 countries we have strong long standing relationships with customers who know us for our powerful combination of leading product brands and applications...

Отзывы: