wtw Stirrox G Скачать руководство пользователя страница 15

StirrOx G

What to do if ...

15

ba25310d07

08/2018

disposed of with the household refuse.

Electrolyte

solution

Disposal according to the safety data sheet.

6

What to do if ...

7

Technical data

Error symptom

Cause

Remedy

The sensor is in the air 
and the display shows 
0.0 mg/l or 0 % O

2

– No connection between 

measuring instrument and sensor

– No electrolyte in the membrane 

head

– Cable defective

– Check connection between 

measuring instrument and sensor

– Replace and refill the membrane 

cap (see section 5.3)

– Return the sensor

The sensor cannot be 
calibrated

– Contaminated membrane head
– Electrolyte depleted

– Replace and refill the membrane 

cap (see section 5.3) 
Subsequently, wait for 30 to 
50 min and recalibrate.

The sensor still cannot 
be calibrated after 
changing the electrolyte 
and membrane cap

– Contaminated electrodes or 

sensor toxification

– Clean the electrodes (see 

section 5.4)

Leak display appears

– Membrane cap not screwed on 

tight enough

– Hole in the membrane

– Screw membrane head tighter
– Replace and refill the membrane 

cap (see section 5.3)

Incorrect temperature 
display

– Temperature sensor defective

– Return the sensor

Mechanical damage to 
the sensor

– Return the sensor

Meter displays 

OFL

 

(display range exceeded)

– Short-circuit between working 

electrode and counter electrode

– Clean the electrodes (see 

section 5.4). If the meter still 
displays 

OFL

, return the sensor.

Stirrer does not move

– No power supply

– Stirring system defective

– Connect power supply (small 

plug)

– Return the sensor

General features

Measuring principle

Membrane covered galvanic sensor

Temperature 
compensation

IMT compensation 
(calculated by the meter)

Temperature sensor

Integrated NTC 30 (30 k

at 25 °C / 77 °F)

Содержание Stirrox G

Страница 1: ...OPERATING MANUAL ba25310e07 08 2018 StirrOx G DISSOLVED OXYGEN SENSOR WITH INTEGRATED STIRRER OPERATING MANUAL...

Страница 2: ...2 ba25310e07 08 2018 StirrOx G Copyright 2018 Xylem Analytics Germany GmbH Printed in Germany...

Страница 3: ...7 4 2 Measuring 7 4 3 Storage 8 5 Maintenance cleaning replacement 8 5 1 General maintenance instructions 8 5 2 Outside cleaning 9 5 3 Changing the electrolyte and membrane cap 9 5 4 Cleaning the elec...

Страница 4: ...Contents StirrOx G 4 ba25310d07 08 2018...

Страница 5: ...ottles 1 Shaft housing 2 Connection head 3 On Off key for stirrer 4 Cable screw joint 5 Stirring paddle 6 WP ST membrane cap filled with electrolyte solution 7 Gold working electrode cathode 8 Insulat...

Страница 6: ...to the instrument or the environment 3 Commissioning Scope of delivery D O sensor StirrOx G filled with electrolyte solution and ready to operate Calibration and storage vessel OxiCal ST ZBK ST acces...

Страница 7: ...sensor is equipped with an integrated stirring system for the incident flow of the membrane Pressing the On Off key switches the stirrer on Pressing once more switches the stirrer off again Measuremen...

Страница 8: ...ables you to place the calibration and storage vessel with the sensor in an upright tilt resistant position on a table 5 Maintenance cleaning replacement 5 1 General maintenance instructions For your...

Страница 9: ...thoroughly rinse with deionized water and recalibrate if necessary 5 3 Changing the electrolyte and membrane cap CAUTION Before starting to work with the sensor please note the GENERAL MAINTENANCE IN...

Страница 10: ...brane head and electrolyte solution see section 5 6 Rinse the sensor head with deionized water Carefully rub and dry the counter electrode with a lint free paper towel Immerse the sensor head includin...

Страница 11: ...crew on the membrane head fingertight using a paper towel Excess electrolyte solution is forced out of the ventilation area Check the filling Inspect the face surface No air bubbles may be present wit...

Страница 12: ...rew off the membrane head Caution Electrolyte solution For disposal of the membrane head and electrolyte solution see section 5 6 Rinse the sensor head with deionized water Using the rough side of the...

Страница 13: ...the membrane cap Additionally you can prevent air bubbles by throwing the first filling away and refilling the membrane cap Rinse the sensor head with electrolyte solution Hold the sensor inclined and...

Страница 14: ...on after 2 minutes 5 6 Disposal CAUTION The ELY G electrolyte solution irritates the eyes and skin When dealing with the ELY G electrolyte solution observe the following points During working activiti...

Страница 15: ...ated after changing the electrolyte and membrane cap Contaminated electrodes or sensor toxification Clean the electrodes see section 5 4 Leak display appears Membrane cap not screwed on tight enough H...

Страница 16: ...r PEEK Thermistor housing VA steel 1 4571 Seals FPM Viton Stirring paddle VA steel 1 4571 Connection cable Lengths 2 m Diameter 6 mm Smallest allowed bend radius Fixed installation 50 mm Flexible use...

Страница 17: ...mended storing method in the calibration and storage vessel OxiCal ST moist Storage temperature 0 50 C 32 122 F Characteristics when delivered Zero signal 0 1 of the saturation value Response time of...

Страница 18: ...assignment Ue V I0 mA I1 mA I2 mA 5 0 4 0 38 189 6 0 4 8 35 147 7 0 5 8 32 137 8 0 6 7 29 121 9 0 7 7 28 108 10 0 8 6 27 100 11 0 9 5 27 95 12 0 10 4 27 89 AE GE HE 7 5 3 1 8 6 4 2 NTC 1 30 kW NTC 2 3...

Страница 19: ...electrode RL G 205 204 Polishing film SF 300 203 680 Accessory kit comprising 3 exchange membrane caps WP ST Electrolyte solution ELY G Cleaning solution RL G Polishing strip SF 300 ZBK ST 202 710 St...

Страница 20: ...Wear parts and accessories StirrOx G 20 ba25310d07 08 2018...

Страница 21: ......

Страница 22: ...buildings factories and farms In more than 150 countries we have strong long standing relationships with customers who know us for our powerful combination of leading product brands and applications...

Отзывы: