Zero adjustment
12
ba77048e01
02/2014
Only perform the zero adjustment against
distilled water in an optically perfect cell.
– Press
1
– In the
setup
menu, call up the
meter setup
submenu.
The following display appears:
– Call up the
zero adjustment
submenu with
3
or
28
.
– Insert a cell with distilled water.
The message,
measuring...
, appears on the display.
Successful zero adjustment for the 10 mm rectangular
cell.
The zero adjustment must be performed
separately for each cell type.
After approx.
2 s: